RLGD-100 REMOTE LASER GAS DETECTOR
23
6
Maintenance & Troubleshooting
The advanced design of the RLGD-100 makes it one of the most reliable leak survey instruments available.
However, should you experience problems with the instrument or suspect that the instrument is not operating
properly, do not use the instrument for leak survey work until the problem is resolved.
Only a qualified RLGD-100 repair technician should attempt to repair or adjust the instrument. There are no
user serviceable components within the instrument that can be repaired or replaced.
6.1.
General Maintenance
Preventive maintenance of the RLGD-100 is limited to:
Please use soft wet cloth to clean dirt or contaminates from the lens, no organic solvent should be used for
cleaning.
Please always keep the lens covered when the instrument is not in use.
Please remove the battery when the instrument is not in use, and store the battery in a proper manner.
Do not point the laser directly to sunlight to avoid possible damage to internal components.
6.2.
Troubleshooting
The following table provides a list of common problems and solution. Should you have a problem not listed
or the solution doesn’t work, please contact FPI for further assistance.
SYMPTOM
SOLUTION
Unable to power on
Try to press again the detection switch about 3 seconds till heard of
the beep tone.
Check if out of battery? (refer to section 3.3 ‘Battery Charging’ in this
manual. )
No detection tone
Check if the sound volume is on? (refer to section 5
.7 ‘Sound & Light’
in this manual.)
No alarm sound
Check if the alarm limit is too high? (refer to section 5
.5 ‘Alarm Level’
in this manual.)
Check if the sound volume is on? (refer to section 5
.7 ‘Sound & Light’
in this manual.)
Unable to charge the
battery
Check if the battery laid in right polarity? Check if the charger,
adapter and power supply properly connected? Check if the charger
red indicator is on? (refer to section 3.3 ‘Battery Charging’ in this
manual.)
Содержание RLGD-100