7
• Shaking suddenly stops.
- Digital displays “Er 01”, indicate the failure of “motor blocked”.
- Digital displays “Er 02”, indicate the failure of “motor
overspeed”
- Pressing the speed control knob or timer control knob first to
stop shaking , and then restart shaking with less load.If
shaking cannot start ,then repeat the process and gradually
reduce load.
6. Maintenance and Cleaning
• Proper maintenance can keep the instrume nt working properly
and lengthen its lifetime.
• Do not spray cleanser into the instrument when cleaning.
• Do not remove the power line when cleaning. Only use
recommended cleansers:
• Before using other method for cleaning or decontamination,
the user must ascertain with the manufacturer that this method
will not damage the instrument. Wear the proper protective
gloves during cleaning of the instrument.
Note:
• Electronic device cannot be clean with cleans er.
• If maintenance service required, the instrument
• must be cleaned first to avoid pollution of
• hazardous substances.
Dyes
Construction materials
Cosmetics
Foodstuffs
Fuels
Isopropyl alcohol
Water containing tenside / Isopropyl alcohol
Water containing tenside / Isopropyl alcohol
Water containing tenside
Water containing tenside