Troubleshooting
The reason for a “Lo” readout
Advice
Read the temperature
Make sure there is no obstacle to the forehead.
when there is hair or sweat
Cool air blows on the forehead
Make sure that there is no cold air
directly blows on the forehead.
The forehead has just been cooled
After the cold compress, wait for 10 minutes
and then measure the measurement
Distance is too far
The measurement distance is recommended to
be 1 - 2 inches (3 - 5cm)
Display Information
Possible Cause
There was no reaction after pressing
The battery is out of power, please replace
the measurement button
the battery; Or the battery is not inserted
into the correct place
low voltage display, please replace the battery
EEPROM abnormal, please send to the
nearest service provider for maintenance
the sensor is not welded well, please send to the
nearest service provider for maintenance
the measurement result over 42.2°C (108°F)
The measured result was below 32°C (89.6°F)
The ambient temperature is too high
The ambient temperature is too low
Specifications
1. Shock-proof type: internal power equipment.
2. According to the degree of shock prevention: BF type application part
3. According to the degree of protection against harmful liquid intake: common equipment.
4. According to the safety level of flammable anesthetic gas mixed with air or mixed with oxygen or
nitrous oxide: equipment that cannot be used in the presence of flammable anesthetic gas.
5. Operating mode: continuous running mode.
6. The thermometer does not have the application part of protection against defibrillation discharge
effect.
7. The thermometer has no signal input and signal outlet.
8. Rated voltage of the equipment: DC 3V.
9. Non-permanent installation of equipment.
10. Electromagnetic compatibility: GB 4824 classification; Group I, Class B equipment.