Instruction Manual
Aquacon 4000
30
6.6
Current Output 1 sub-function
In this subfunction you set the current output range of the transmitter for conductivity values.
ENT
SETUP
HOLD
4
SETUP
HOLD
4
ENT
ENT
SETUP
HOLD
4
ENT
SETUP
SETUP
HOLD
HOLD
µS
µS
4
4
SETUP
SETUP
HOLD
HOLD
4
4
ENT
ENT
ENT
ENT
SETUP
SETUP
HOLD
HOLD
µS
µS
4
4
1.
Select the “CUR.1” subfunction, then
press the ENT key
.
2.
Selecting the output type: press the
▲
or
▼
key
to select the desired output type: 0-
20 mA or 4-20 mA.
Press the ENT key
to confirm your selection.
3.
Setting the conductivity value at which the transmitter output will be 4 mA (or 0 mA):
press the
▲
or
▼
key
to set the conductivity value to be equivalent to 4 mA or 0 mA,
respectively.
Press the ENT key
to confirm your setting.
4.
Setting the conductivity value at which the transmitter output will be 20 mA: press
the
▲
or
▼
key
to set the conductivity value to be equivalent to 20 mA. Press the ENT
key to confirm your setting.
5.
Selecting out of range current: Press the
▲
or
▼
key
to switch “CU22” on or off. If
“CU22” is on, the output signal will jump to 22 mA when the measured conductivity value
is below or above the values set in points 3 & 4 above.
Press the ENT key
to confirm
your setting.
Note
: the “CU22” setting (on/off) of current output 1 applies to current output 2 as
well.
Continue with Setup mode procedures, or return to Measurement mode by pressing the
▲
and
▼
keys simultaneously (escape).
Содержание Aquacon 4000
Страница 1: ...Instruction Manual Aquacon 4000 Conductivity Controller Transmitter 01 2004...
Страница 10: ...Instruction Manual Aquacon 4000 5 2 3 Unit overview Wall mounting version Panel mounting version...
Страница 11: ...Instruction Manual Aquacon 4000 6...
Страница 15: ...Instruction Manual Aquacon 4000 10...
Страница 27: ...Instruction Manual Aquacon 4000 22...
Страница 45: ...Instruction Manual Aquacon 4000 40...
Страница 51: ...Instruction Manual Aquacon 4000 46 Panel mounting version view from top...
Страница 58: ......