55
SPECIFICATIONS
SYSTEM PERFORMANCE
Power Source
Four 1.5V AA alkaline batteries
Size of Monitor w/o Cuff
155 (L) x 110 mm (W) x 69 mm (H),
340g with batteries
Memory
Maximum 450 memory records
Power Saving
Automatic power off if system idle for 3
minutes (normal mode) or 5 minutes (RF
mode).
System Operating
Condition
50
°F to
104
°F
(10
°C to
40
°C
), below 85%
RH
Monitor Storage Condition
-4
°F to
140
°F
(-20
°C to
60
°C
), below 95%
RH
Power Supply Input
DC +6V / 1A (max) via Power Plug
BLOOD GLUCOSE MEASUREMENT PERFORMANCE
Measurement Unit
mg/dL
Linear Range
20 to 600 mg/dL (1.1 to 33.3mmol/L)
Precision
±5 % (CV)
Accuracy
±15 mg/dL (0.83 mmol/L) when glucose <
75mg/dL (4.2 mmol/L)
±20% when glucose ≥ 75mg/dL (4.2
mmol/L)
Ketone Warning
glucose value is over 240 mg/dL (13.3
mmol/L)