   INTRODUCTION

Thank you for choosing 

FORA IR18

 ear thermometer. This 

innovative medical device relies on advanced infrared (IR) 
technology to measure temperature instantly.

   HOW DOES IT WORK

The thermometer measures the infrared heat generated by 
the eardrum and its surrounding tissue. The thermometer 
then converts it into a temperature value shown on the LCD 
screen.

   IMPORTANT SAFETY INSTRUCTIONS

READ THIS BEFORE USE. KEEP THESE INSTRUCTIONS IN A 
SAFE PLACE.

The following basic safety precautions should always be 
taken.
1. Close supervision is necessary when the thermometer is  
  used by, on, or near children, handicapped persons or  
 invalids.
2. Use the thermometer only for the intended use described  
  in this manual.
3. Do not use the thermometer if it is not working properly,  
  or if it has suffered any damage. 
4. Do not use accessories which are not supplied or  

 

  recommended by the manufacturer.
5. Proper maintenance is essential to the longevity of your  
  device. If you are concerned about your accuracy of    
  measurement, please contact the local customer service  
  or place of purchase for help.

   CAUTIONS AND WARNINGS

 ● 

As with any thermometer, proper technique is crucial to  

  getting accurate temperature readings. Please read this  
  manual thoroughly and carefully before use.

 ● 

Always operate the thermometer in temperatures  

 

  between 50°F to 104°F (10°C to 40°C) and relative   
  humidity less than 95%.

 ● 

Always store the thermometer in a cool and dry place:  

  temperatures between -4°F to 140°F (-20°C to 60°C);  
  relative humidity less than 95%. Avoid direct sunlight.

 ● 

Avoid dropping the thermometer.

 ● 

Use only 

FORA IR18

 probe covers for the infrared ear   

 thermometer.

 ● 

For proper hygiene, do not share or reuse probe covers.  

  Damaged probe covers may result in an error display.

 ● 

Do not touch the probe cover tip. 

 ● 

Do not disassemble the thermometer.

 ● 

Basic safety precautions should always be observed,   

  especially when the thermometer is used on or near    
  children and disabled persons.

 ● 

This thermometer is not intended to be a substitution for  

  a consultation with your physician.

 ● 

Keep probe covers out of reach of children.

 ● 

Temperature measurements may differ from left to right  

  ear. Always measure by using the same ear.

   RESTRICTIONS OF USE

This thermometer is clinically proven to produce accurate 
temperature measurements. However, please be advised if 
you have the following situations:

 ● The accuracy cannot be ensured for a person who has a  

  deformity in their ear, which prevents the thermometer  
  probe from properly being inserted into the ear canal.

 ● The accuracy cannot be ensured when blood or drainage  

  is found in the ear canal.

 ● If ear drops or medications have been placed in one ear,  

  take measurements from the other ear.

 ● For a person who wears an ear plug or hearing aid,    

  remove the device and wait for 15 minutes before taking  
 temperature.

NOTE:

 Never try to clean inside your ears. You may 

accidentally damage the eardrum or its surrounding tissues. 
Remove excess earwax only when you can reach it with a 
clean cloth. Consult a physician if you suspect the presence 
of excess earwax.

   APPEARANCE AND KEY FUNCTIONS OF THE
   THERMOMETER 

Scan Button
Probe
LCD Screen

1

2

3

On/Memory Button
Battery Cover
Probe Cover

4

5

6

1

2

6

3

4

5

   HINTS ON TAKING TEMPERATURES

1. 

As with other thermometers, you may observe slight 

variations in consecutive measurements. 

It is  

 

recommended that you take 3 temperature readings and 
use the highest one for the following situations:

 ● Infants younger than 3 months old.
 ● Children younger than 3 years old and who have a    

  compromised immune system and the presence /    
  absence of fever is critical.

 ● When you are learning to use the thermometer.

2. 

Do not take a reading while the patient is moving and/or 

talking. Wait for 15 minutes after any of the following 
situations before taking a measurement:

 ● When the ear has been covered.
 ● After exercising, swimming, or taking a bath.
 ● When the patient has been exposed to extreme  

 

 temperature.

3. 

To take accurate readings, the ear must be free from 

excessive earwax build-up.

   USING THE DEVICE

Replacing the Probe Cover
1.

 Check if the probe cover is clean and new. 

2.

 Secure the cover to the probe.

NOTE: 

 ● 

The used probe cover should be removed after each    

  measurement to ensure accuracy. 

 ● 

To ensure proper hygiene, attach a new probe cover each  

  time and do not touch its tip.

2

1

   INTENDED USE

FORA IR18

 ear thermometer is intended for the intermittent 

measurement and monitoring of human body temperature 
from the ear canal. The device is intended for home use by 
people of all ages.

   LCD SCREEN 

3

2

4

5

6

7

1

   RESULT INDICATORS

“Lo” 
Result with green 
backlight
Result with red 
backlight
“Hi” 

Temperature ≤ 89.5°F (31.9°C)
89.6°F (32°C) ≤ Temperature

< 100.3°F (37.9°C)
1

00.4°F (38°C) ≤ Temperature

< 109.4°F (43°C)

Temperature ≥ 109.5°F (43.1°C) 

5

1.

 Press and release the On/Memory button to turn on the 

thermometer. When ready, the thermometer  displays the 
last measurement.

2.

 Gently fit the probe with a clean probe cover into the ear 

canal.

3.

 Press and release the Scan button. You will hear a beep 

to indicate the measurement is complete and the 
thermometer can be removed. 

4.

 Read the result. “     ” is shown together with a 

temperature value.

5.

 Discard the used probe cover into a trash can.

NOTE: 

Turn off the thermometer by pressing On/Memory 

button twice. It will automatically turn off if left idle for 30 
seconds.

   MEASURING TEMPERATURE

2

3

11

14