EN-20
Manufacturer’s declaration-electromagnetic emissions
The FocusTemp is intended for use in the electromagnetic environment (for
home healthcare and professional healthcare) specified below.
The customer or the user of the FocusTemp should assure that it is used in
such an environment.
Emission test
Compliance Electromagnetic
environment-guidance
(for home healthcare and
professional healthcare)
RF-emissies CISPR 11
Group 1
The FocusTemp uses RF
energy only for internal use.
Therefore, its RF emissions
are very low and are not likely
to cause any interference
from nearby electronic
equipment.
RF-emissies CISPR 11
Class B
The FocusTemp is suitable
for use in all establishments,
including domestic
establishments and those
directly connected to the
public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations
/flicker emissions IEC 61000-
3-3
Not applicable
Manufacturer’s declaration-electromagnetic immunity
The FocusTemp is intended for use in the electromagnetic environment (for
home healthcare and professional healthcare) specified below.
The customer or the user of the FocusTemp should assure that it is used in
the environment specified below.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment-
guidance (for
home healthcare
and professional
healthcare
environment)
Electrostatic
discharge(ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Содержание FocusTemp