Replacement of parts subject to wear and tear
Check the following parts subject to wear and tear and replace where
applicable:
● Light guides (change after 2,000 sterilization cycles or every two
years)
● Silicone key pad of the handpiece (change after 400 treatments/
sterilizations)
● Fiber cutter (change after 400 treatments/sterilizations or every two
years)
● Rechargeable battery (change after 1000 charging cycles or every
two years)
For further informations, see chapter "Cleaning, disinfection and
CAUTION
Only use orignal parts from the manufacturer, see "Spare parts [ → 27]".