Operating Instructions
båÖäáëÜ
cê~å´~áë
7 Cleaning, disinfection and sterilization
Following treatment, switch off the Aquamarine and disconnect the power
cable from the power supply.
Control unit, handpiece body and handpiece are supposed to be cleaned
and wipe-disinfected.
Dispose the single-use fiber tips.
The removable handpiece sleeve, light guides, fibercutter and bending
tool are supposed to be cleaned and sterilized.
For the number of sterilization cycles, see chapter "Replacement of parts
subject to wear and tear".
7.1
Cleaning
Handpiece sleeve
1. Remove the handpiece sleeve from the handpiece body by gently
pulling the sleeve off the handpiece body after disposing the single-
use fiber tip or the light guide.
2. Clean the handpiece sleeve with a suitable brush under running
water.
Light guide (ActiveTip)
➢ Clean the light guide under running water (< 38 °C, water must be at
least drinking water quality).
NOTICE
When following these instructions wear gloves.
CAUTION
Do not clean and disinfect the parts using a washer-disinfector! The
parts may be seriously damaged.
NOTICE
Manual cleaning must always be combined with disinfection.
CAUTION
Danger of damage to the optical system
Reattach the protective cap to the optical system of the handpiece
immediately after dismounting the single-use fiber tip or therapy light
guide. Do this before taking any hygienic measures.
NOTICE
Never clean in an ultrasonic bath!