background image

Controller operation

In normal operation the controller displays cabinet temperatures in 
°C (default) or user-selected °F. °C temperatures are displayed to 1 
decimal point. 

Rocker buttons to the right of the temperature display control all 
programming functions.

The controller is pre-programmed with a +3.5°C (+38°F) set point 
which provides a compressor cut-in at +3.5° (+38°F) and cut-out at
 +5.0°C (+41°F). 

All set points have a 1.5°C differential. The 1.5°C differential means 
that with a +3.5°C set point, for example, the compressor will turn off 
at +3.5°C (+38°F) and turn on when the cabinet temperature reaches 
+5°C (+41°F).

This temperature may not reflect the temperature of the blood products in the refrigerator. The temperature of the 
blood products will be reflected on the recorder chart and the recorder's digital temperature display. At times there 
may be a difference between the two displayed temperatures. Adjust the controller SET value to achieve your 
desired blood product temperature as displayed on your chart recorder.

5

2.5

°C

SET

prg

refrigeration
system
energized

Fig. 6

To display temperature cut-out

STEP INPUT 

DISPLAY

   1 

Press and release SET 

Current cut-out temperature will display for approximately 5 seconds.

 

 

Display will return to current refrigerator temperature.

To change temperature cut-out 

STEP INPUT 

DISPLAY

   1 

Press and hold SET for 

Current cut-out temperature displayed, refrigeration system

 

3 seconds 

energized LED flashes

   2 

Press UP or DOWN arrows to 

New cut-out temperature displayed

 

desired cut-out temperature

   3 

Press and release SET 

New cut-out temperature blinks three times, then current refrigerator 

 

 

temperature will display 

Refer to the chart below for the set point for your application's required temperature range.

SET POINT °C 

CUT-IN °C 

CUT-OUT °C 

NORMAL DISPLAY RANGE °C

     +3.5°C 

+5.0°C 

+3.5°C 

1.5° to 5.0°

Содержание REF5-BB

Страница 1: ...od Bank Refrigerator 00161943R00 801 Church Lane Easton PA 18040 USA Toll free 800 523 9361 610 252 7301 Fax 610 250 0696 www follettice com Following installation please forward this manual to the ap...

Страница 2: ...we can help in the filing of a claim if necessary Specifications REF5 BB 34 00 height for freestanding use or installation below standard 36 915mm high counter 2 drawer 4 8 cu ft capacity up to 50 45...

Страница 3: ...module when ordered is packed in the lower drawer Use the stainless steel sheathed probe included in the module package in conjuction with the product simulating bottle mounted inside the refrigerato...

Страница 4: ...Fig 4 2 4 Cover hinge screw holes with screw hole plugs removed from opposite side 5 Reverse door and reinstall hinge screws 6 Reinstall latch on opposite side 7 Remove screws and handle from door Fi...

Страница 5: ...re display At times there may be a difference between the two displayed temperatures Adjust the controller SET value to achieve your desired blood product temperature as displayed on your chart record...

Страница 6: ...rogramming of factory settings Programming key part number is 00161950 Programming refrigerator from the programming key download 1 Turn OFF refrigerator 2 Remove 6 screws from front panel to access b...

Страница 7: ...5 for changing the temperature set point The 1 5 C cut out differential will be maintained regardless of the controller set point Defrosting REF5 BB Series undercounter refrigerators do not require m...

Страница 8: ...tray allows technicians to partially slide the condensing unit from the refrigerator back without cutting refrigerant lines 1 Remove rear panel Fig 10 1 2 Remove two bolts securing condensing unit to...

Страница 9: ...as a guide remove wires one at a time from back of existing controller and connect to corresponding terminals on replacement controller 7 Gently remove faceplate from existing controller and install o...

Страница 10: ...ures 1 Remove the rear access panel Fig 10 2 Turn the power switch to the on position 3 Following the instructions on page 5 verify that the temperature controller is set to the original factory set p...

Страница 11: ...ller contact terminals for power Replace controller if needed 5 Replace controller and or probe 1 Allow to cool or replace 2 Replace as required 3 Replace compressor 1 Replace fan motor 1 Replace cont...

Страница 12: ...0155564 2 Fan guard 00152892 3 Fan blade 00152991 4 Bracket fan motor 00152983 5 Fan motor evaporator 00104919 6 Air baffle 00165126 7 Drain pan evaporator 00162511 8 Evaporator 00151563 9 Expansion v...

Страница 13: ...ser 00157347 3 Condenser fan blade 00105007 4 Condenser fan motor 00104992 5 Fan motor bracket 00157412 6 Compressor 00104950 7 Starting capacitor 00104968 8 Filter drier 502724 Not shown Cap starting...

Страница 14: ...6 Gasket REF5 BB 00125732 7 Strip sealer set of 4 REF5 BB 00130138 8 Drawer 00165134 9 Drawer slides Reference only 10 Nut acorn each 4 required per drawer 00161802 11 Screw 5 8 each 4 required per dr...

Страница 15: ...components Reference Description Part 1 Temperature controller 00161570 2 Faceplate C 00157644 Not shown Probe harness controller 00165167 3 Power switch 00114371 4 Front panel includes 00114371 and 0...

Страница 16: ...00161943R00 01 07 801 Church Lane Easton PA 18040 USA Toll free 800 523 9361 610 252 7301 Fax 610 250 0696 www follettice com...

Отзывы: