Control Panel Functionality Check
The first time the unit is powered on, follow the operation below to ensure control panel functionality.
Manual check of Control Panel
Operation
Key operated
Control panel description
1
Connect to the power
source.
-----
Five seconds after turning on the unit, the
temperature inside the unit should appear on
temperature display (LED Screen).
2
Press “Alarm Test” key.
ALARM TEST
Two seconds after pressing the button, the buzzer
should sound every 1 second. Let the buzzer sound
at least 3 times.
3
Press “Alarm Test” key.
ALARM TEST
Within 2 seconds of pressing the button, the buzzer
should stop. (The buzzer should stop sounding after
two second of pressing the button).
4
Press “SET” key.
SET
Within 2 seconds of pressing the button, you will
hear a beep once. The display will show “5.0” on
LED Screen and it will start blinking.
5
Press
▼
key.
▼
Within 2 seconds of pressing the button, you will
hear a beep once. The display will show “4.9” on
LED Screen and it will start blinking.
6
Press
▼
key two times.
▼
Within 2 seconds of pressing the button, you will
hear a beep once. The display will show “4.7” on
LED Screen and it will start blinking.
7
Press
▲
key three times.
▲
Within 2 seconds of pressing the button, you will
hear a beep once. The display will show “5.0” on
LED Screen and it will start blinking.
9
Open the door
-----
Two seconds after opening the door, the door check
lamp will turn red.
10
Close the door
-----
Two seconds after closing the door, the door check
lamp will turn off.
11
Open the door
-----
Two seconds after opening the door, the door check
lamp will turn red.
12
Keep the door open
-----
After 120 seconds, the alarm LED will turn red and
will start beeping every second. Let the unit beep at
least 3 times.
13
Close the door
-----
Two seconds after closing the door, the door check
LED will turn off.
8
REFNAT5-SD, REFNAT5-GD