NOTICE!
Wiring diagrams, schematics, and other
documents which are needed for repairing the unit, will be
supplied by KaVo Dental on request.
MEASUREMENT CRITERIA FOR LOADING FACTOR
CONDITIONS
Exposure time
Exposure time consists of beginning and ending points as
measured by a calibrated x-ray monitor at 70% of the peak
radiation waveform.
kVp
The high voltage peak value measured over the high
voltage feedback resistor with a calibrated voltage device.
mA
The tube current mean value calculated by dividing the
voltage over the feedback resistor value. The voltage is
measured with a calibrated voltage device.
The nominal x-ray voltage 70kV is obtained at highest
tube current 7mA.
The nominal tube current 7mA is obtained at the
highest tube voltage 70kV.
The max. electric output is obtained at 70 kV tube
voltage and 7 mA tube current.
The nominal power/exposure: 490 W
Instrumentarium Dental reserves the right to make
technical changes at any time.
Содержание 214760
Страница 1: ...FOCUS Intraoral X ray User Manual ENGLISH 214760 rev 1 0 805 4936...
Страница 2: ......
Страница 4: ......
Страница 6: ...ii KaVo Dental GmbH 214760 rev 1...
Страница 21: ...2 Unit description 214760 rev 1 KaVo Dental GmbH 5 Exposure time control...
Страница 22: ...2 Unit description 6 KaVo Dental GmbH 214760 rev 1...
Страница 32: ...5 Error messages 16 KaVo Dental GmbH 214760 rev 1...
Страница 52: ...9 Recommended exposure times 36 KaVo Dental GmbH 214760 rev 1...