Operation Manual RTM MP Plus
06.03.2019
19
The keys on the operating panel of the EMGZ482R.MP version are all deactivated and
have not their normal function as in the standard version.
Key
Key function during operation
Key deactivated
Key deactivated
Key deactivated
Key deactivated
Keys deactivated
Key deactivated
Key deactivated
5.3
Offset Compensation Procedure
With the Offset Compensation one can compensate the roller weight of the sensor.
This procedure is always performed before the calibration. The force measuring roller
should not be loaded while the Offset Compensation is being done. The roller weight
compensation is performed for both channels (sensor 1 and sensor 2).
Procedure:
1.
Remove any wires that may apply a force to the roller.
2.
Close a contact between terminal 10 and 13 on the EMGZ482R.MP (see Fig. 6).
The switch time must be >100 ms.
3.
This connection will perform the offset compensation on both force measuring
rollers (sensor 1 and 2).