Operating Manual CMGZ433
33
Offset feedback
Use:
This parameter stores the value determined with
Find offset
in [Digit].
It is not necessary to note this parameter, because a new offset
adjustment is done very easy; also when changing the whole
electronic unit.
The offset can also be inputted by using the
↑
↓
←
keys.
Range:
-4000
to
4000
Default:
0
Increment:
1
Unit:
[Digit]
Gain feedback
Use:
This parameter stores the value determined with
Calibration feedback
,
resp. you can input a value calculated using the formulas described
under „9.4 Calibrating the measuring amplifier“, if the
material tension
cannot be simulated.
Range:
0.100
to
9.000
Default:
1.000
Increment:
0.001
Unit:
[-]
Nominal force of sensor
Use:
To get the correct force value,, the electronic unit has to know the
nominal force of the sensors.
Range:
1
to
9999
Default:
1000
Increment:
1
Unit:
[N,kN]
Unit of force
Use:
This parameter stores the force unit of the sensor.
Range:
N,
kN
Default:
N
Sensitivity of sensor
Use:
To get the correct force value,, the electronic unit has to know the
sensitivity of the force sensors, that means how much signal the sensor
will answer by nominal force. Standard for FMS force sensors is
1.8mV/V.
Range:
0.1
to
3.0
Default:
1.8
Increment:
0.1
Unit:
[mV/V]