background image

CAUTION:

DO NOT ATTEMPT TO LOOSEN "LOCK NUT", THIS WILL RESULT IN   

BREAKING THE ADJUSTMENT SCREW.

STH PUMPS

STQ PUMPS

ADJUSTMENT
SCREW

LOCK ADJUSTMENT

(REMOVABLE THREAD LOCK)

ADJUSTMENT

SCREW

HEX LOCK NUT

DO NOT

LOOSEN

3.4

Dispense Volume Values for FMI 

STH/STQ Pumps:

NOTE

: To assure the best performance set the pump

displacement volume as large as possible.  

FMI pumps

are calibrated at the factory to the maximum value.

Both STH and STQ pumps can be adjusted down to as
little as 10% of the rated flow.

3.5

Adjusting Dispense Volumes:

To adjust dispense volumes rotate the stroke-length 
adjustment screw using the appropriate hex driver 
(see Required Tools on page 1). To increase the flow, 
turn the screw counter-clockwise. To decrease the 
flow, turn the screw clockwise 

CCW=Increase, CW=Decrease

CAUTION: DO NOT LOOSEN THE HEX LOCK 
NUT WHEN   ADJUSTING THE STROKE RATE.  IT IS 
SET AT THE FACTORY AND SHOULD NOT  BE 
TAMPERED WITH. See Figure E

3.6

Lock setting: 

Once the pump is calibrated apply 

a drop of removable wicking thread lock between the 
lower pivot pin and the adjustment screw threads. 

See  Figure E.

3.7 Dispensing Hints:

1)  Speed

- Optimum results for fluids 500 cps or 

less is between 120-350 rpm.

2)  Cavitation:

Use the largest suction tubing you

can to avoid cavitation.

3)  Splashing

can usually be avoided by 

modifying dispense tip to larger I.D.and/or
decreasing dispense speed.

4)  Hanging drop at dispense tip can be avoided by:

a) Use of rigid discharge tubing
b) Small dispense tip
c) Increasing speed

5)  High viscosity dispensing requires:

a)  Large suction tubing
b)  Pressurizing suction reservoir

6)  Bubbles in discharge

a)  Suction fitting leak
b)  Cavitation

7)  Need Help?  Call, fax, or email us… We can 

help solve almost any fluid control problem.

3.8 SPECIAL  MOTORS AND CONTROLLERS
FMI offers a complete variety of stepper motors
and stepper motor controllers from simple quick 
start control to complex application specific stepper 
motor control.  Our standard step motor controllers 
are:

SCST

General Purpose Stepper Motor Control for

quick start control of FMI STH and STQ pumps.

ICST

Intelligent Stepper Motor Controller 

which includes an embedded  microprocessor for 
custom programming of FMI stepper motor pumps
to meet specific application operations.

IDS 2000

Industrial Dispenser/Pump which includes 

the SCST and FMI pump head integrally mounted 
in a rugged stainless steel enclosure.  

STH00

0.002

0.025

STH0

0.005

0.050

STH1

0.010

0.100

STQ0

0.008

0.080

STQ1

0.003

0.032

STQ2

0.007

0.072

STQ3

0.128

1.280

NOTE:FOR ADDITIONAL REFERENCES  SEE  H431 AND Q431

Figure E

3

MODEL

ML/REV 

MIN

ML/REV

MAX

5 Aerial Way, Suite 500, Syosset NY 11791 Tel: 516-922-6050 Fax: 516-624-8261 

Website: www.fmipump.com   Email: [email protected]

ST431-02

Содержание STH00

Страница 1: ...s or liability for any misuse that will cause damage to product and or injury to the individual Power cables and leads should not be bent pulled or inserted by excessive force Otherwise there is a threat of electrical shock or fire Replace any inline fuses only with fuse rating as specified by Fluid Metering Inc When pump drive is under operation never point discharge tubing into face or touch any...

Страница 2: ...inlet port will be on the left when the motor rotates clockwise see Figure B for the STH pump and Figure C for the STQ pump OUTLET STQ PISTON FLAT SENSOR FLAG RELATIONSHIP TO INLET AND OUTLET PORTS CLOCKWISE MOTOR ROTATION PISTON FLAT PISTON FLAT INLET CLOCKWISE MOTOR ROTATION OUTLET ADJUSTMENT SCREW FLAG SENSOR STH PISTON FLAT SENSOR FLAG RELATIONSHIP TO INLET AND OUTLET PORTS INLET 3 3 Check the...

Страница 3: ...ion 3 Splashing can usually be avoided by modifying dispense tip to larger I D and or decreasing dispense speed 4 Hanging drop at dispense tip can be avoided by a Use of rigid discharge tubing b Small dispense tip c Increasing speed 5 High viscosity dispensing requires a Large suction tubing b Pressurizing suction reservoir 6 Bubbles in discharge a Suction fitting leak b Cavitation 7 Need Help Cal...

Страница 4: ...SERIES BIPOLAR 4 LEAD LEAD COLOR CONNECTION DRIVER PIN CONNECTION 1 8 DEGREES OF ROTATION 2 1 1 3 2 3 2 2 5 5 DEGREES C PER WATT 0 0017 0 012 OZ IN SEC KG M 10 4 0 0 028 OZ_IN Nm OZ_IN Nm 89 0 63 63 0 44 82 0 58 2 3 1 06 0 206 2 5 89 0 63 116 0 82 2 3 0 53 0 206 3 5 89 0 63 116 0 82 9 2 2 12 0 412 1 75 OZ_IN Nm OZ_IN Nm mH OHMS Vpk RAD SEC AMPS UNIPOLAR 6 LEAD PARALLEL SERIES UNITS FMI STEP MOTOR ...

Отзывы: