Primary actions after installation
3 - Fuel computer:
(if installed). If it's installed the fuel flow
transducer, BEFORE rely on informations provided by the fuel
computer section you must:
- Verify that the K-factor set in the Vigilus is pertinent to the
installed fuel flow transducer (for the Flybox® TFTHP is 416400).
- Execute the fuel flow transducer calibration as explained in
chapter 1.4. Without calibration the fuel computer
informations may be wrong, even if the nominal K-factor is
correct for the fuel flow transducer used.
After calibration, the K-factor should have been calculated
automatically and at best for every single installation. You must
still check for some time if the remaining quantity indicated are
reliable compared to the refuelling performed. For example, if
the instrument indicate a remaining quantity of 35 liters and you
know that the tanks capacity is 80 liters, filling the tanks should
require approximately 45 liters; in case of much difference redo
the calibration.
Consider also that, during use, little errors accumulate and if
you never fill the tanks you never “reset” all these errors.
Содержание Vigilus
Страница 1: ...Revision 1 0 18 7 2017 For firmware version 2 14 Flybox Operating manual ...
Страница 2: ...Page intentionally left blank ...
Страница 3: ...SECTIONS INSTRUMENT CONFIGURATION OPERATING INSTRUCTIONS TECHNICAL SPECIFICATIONS ...
Страница 10: ...Page intentionally left blank ...
Страница 89: ...Page intentionally left blank ...
Страница 90: ...Page intentionally left blank ...
Страница 91: ...Page intentionally left blank ...