Versiv Cabling Certification Product Family
Users Manual
180
The SmartLoop Test
The SmartLoop test lets you connect the far ends of the two fibers
in a link and do one OTDR test to get separate results for each
fiber.
You use a launch cord to connect the fibers together at the far
end of the link. When you do the SmartLoop test, the tester uses
the launch compensation settings to remove the effects of the
loopback cord and its connectors.
The Auto and Manual settings for the SmartLoop test operate the
same as for the OTDR test. See Table 7 on page 153.
Figure 72 shows the equipment for the SmartLoop test.
Figure 72. Equipment for SmartLoop Tests
GPU169.EPS
Versiv unit with OTDR module
installed
Three launch cords for launch,
tail, and loopback connections
(match the fiber to be tested)
Short patch cord (30 cm (12 in)
maximum) and adapter for
bidirectional tests
FI-1000 video probe with USB
connector
Fiber cleaning supplies
AC adapter with line cord
(optional)
A
B
E
F
D
C
Содержание Versiv
Страница 38: ...Versiv Cabling Certification Product Family Users Manual 24...
Страница 86: ...Versiv Cabling Certification Product Family Users Manual 72...
Страница 202: ...Versiv Cabling Certification Product Family Users Manual 188...
Страница 214: ...Versiv Cabling Certification Product Family Users Manual 200...
Страница 234: ...Versiv Cabling Certification Product Family Users Manual 220...
Страница 244: ...Versiv Cabling Certification Product Family Users Manual 230...
Страница 256: ...Versiv Cabling Certification Product Family Users Manual 242...
Страница 258: ...Versiv Cabling Certification Product Family Users Manual 244...