Testing for Transducer Leakage Current
11
Testing for Transducer Leakage Current
Note
Perform the Leakage Test only if the Conductivity Test passes.
1.
Select
LEAKAGE
.
2.
Press the
ON/TEST
button to perform a leakage measurement.
The green
PASS
light illuminates if the transducer passed the leakage
current test. The red
FAIL
light illuminates if the transducer failed the
leakage current test. If there is less than 20
µ
A of leakage current, the red
light pulses, indicating a possible open circuit condition with invalid test
results.
Maintenance
Your ULT800 needs little maintenance or special care. However, treat it
as a calibrated measuring instrument. Avoid dropping or other mechanical
abuse that could cause a shift in the calibrated settings.
Cleaning
W
Caution
Do not pour fluid onto the ULT800 surface; fluid seepage
into the electrical circuitry may cause ULT800 failure.
W
Caution
Do not use spray cleaners on the ULT800; such action may
force cleaning fluid into the ULT800 and damage electronic
components.
Clean the ULT800 occasionally utilizing a damp cloth and mild detergent.
Take care to prevent the entrance of liquids.
Wipe down the adapter cables with the same care. Inspect them for
damage to and deterioration of the insulation. Check the connections for
integrity.