Fluke ULT800 Скачать руководство пользователя страница 1

 

 

ULT800

 

TEE Transducer Leakage Current Tester 

Users Guide 

 

 

PN 2461434 
August 2005

 

© 2005 Fluke Corporation, All rights reserved. Printed in USA 
All product names are trademarks of  their respective companies.

 

Содержание ULT800

Страница 1: ...ULT800 TEE Transducer Leakage Current Tester Users Guide PN 2461434 August 2005 2005 Fluke Corporation All rights reserved Printed in USA All product names are trademarks of their respective companies...

Страница 2: ...y This warranty is limited to repairing the instrument to Fluke Bio medical s specifications Warranty Disclaimer Should you elect to have your instrument serviced and or calibrated by some one other t...

Страница 3: ...onductivity of the fluid A special dual element probe also connected to the ULT800 tests the conductivity of the fluid A green PASS light or a red FAIL light indicates the results of the conductivity...

Страница 4: ...in their original condition and contact the carrier immediately to file a claim If the instrument is delivered in good physical condition but does not operate within specifications or if there are any...

Страница 5: ...1 Every product returned to Fluke must have a Return Material Authorization RMA number To obtain an RMA contact Fluke through one of the following methods Phone 888 99FLUKE 888 993 5853 or 425 446 556...

Страница 6: ...R LEAKAGE CURRENT TESTER INSTRUCTIONS 1 Connect the transducer and the adapters to the ULT 800 2 Press the ON TEST button and wait for the READY light 3 Select CONDUCTIVITY and press the ON TEST Butto...

Страница 7: ...est or the Leakage Test fails Pulses red when the Leakage Test results in less than 20 A indicating a possible open circuit condition with invalid test results LOW BAT Flashes red to indicate that the...

Страница 8: ...ments on a single battery Conductivity Limit to pass greater than 250 A 5 Leakage Limits to pass less than 100 A 5 and greater than 20 A 5 A Dimensions 6 5 x 3 7 x 1 5 in 17 x 19 x 4 cm Weight 12 oz 3...

Страница 9: ...nd compares the results to an internal threshold The instrument displays the results as a PASS or FAIL indication It also performs an internal self calibration on each measurement cycle The ULT800 mak...

Страница 10: ...LT 800 2 Press the ON TEST button and wait for the READY light 3 Select CONDUCTIVITY and press the ON TEST Button Wait for the PASS light 4 Select LEAKAGE and press the ON TEST button again observe PA...

Страница 11: ...tton Wait for the PASS light 4 Select LEAKAGE and press the ON TEST button again observe PASS or FAIL light If test fails DO NOT USE TRANSDUCER Refer to the user s manual ON READY PASS FAIL LOW BAT TE...

Страница 12: ...t remains on for 12 seconds The ULT800 then powers off to conserve the battery To resume testing power on the ULT800 allow the self check battery test to complete and then reinitiate the test 1 Select...

Страница 13: ...ance Your ULT800 needs little maintenance or special care However treat it as a calibrated measuring instrument Avoid dropping or other mechanical abuse that could cause a shift in the calibrated sett...

Страница 14: ...h the anti tamper case label broken is out of calibration A ULT800 that is out of calibration can cause excessive leakage current exposure to the patient risk of injury to the patient could result W C...

Страница 15: ...cuson 156 series probes 2392453 600 202FG Ultrasound Transducer Adapter GE YMS RT 2392466 600 203FG Ultrasound Transducer Adapter GE CGR 2392475 600 204FG Ultrasound Transducer Adapter GE LogiQ 223160...

Страница 16: ...s Acuson 260 series probes 2231811 600 360 Ultrasound Transducer Adapter Siemens Acuson 360 series probes 2392591 600 MPFG Ultrasound Transducer Adapter Siemens Acuson MP series probes 2392427 600 102...

Страница 17: ...f electric shock Standby On P Manufacturer s declaration of product compliance with applicable EU directives LISTED 58GB E233218 UL Listing mark Do not mix with solid waste stream Dispose using a qual...

Страница 18: ...ULT800 Users Guide 16...

Отзывы: