Do not touch voltages >30 V ac rms, 42 V ac peak, or
60 V dc.
Do not use the Product around explosive gas, vapor,
or in damp or wet environments.
Do not exceed the Measurement Category (CAT)
rating of the lowest rated individual component of a
Product, probe, or accessory.
Do not use test leads if they are damaged. Examine the
test leads for damaged insulation, exposed metal, or if
the wear indicator (if present) shows. Check test lead
continuity.
Keep fingers behind the finger guards on the probe.
Do not use in CAT III or CAT IV environments without
the protective cap installed on test probe. The
protective cap decreases the exposed probe metal to
<4 mm. This decreases the possibility of arc flash from
short circuits.
1000 V CAT II without protective cap installed.
1000 V CAT III, 600 V CAT IV with protective cap
installed.
Do not use a shrouded banana jack accessory such as
an alligator clip with an unshrouded 4 mm banana
probe tip terminal in a hazardous electrical
installation. Due to the lack of a shroud, the probe tip
connector does not provide suitable insulation to
accessible parts with this type of assembly.