TL2X4W-TWZ English Instruction Sheet 

Page  

 

Safety Information 

XW

 Warning 

To avoid possible electric shock or personal injury: 

 

Do not use the test leads to measure voltages 
above 33 V rms, 46.7 V peak or 70 V dc. 
Voltages above these levels will cause electric 
shock. 

 

Do not use the test leads for measurements 
within Measurement Categories II, III, and IV. 

 

Use the test leads only as specified by Fluke or 
the protection provided may be impaired. 

 

Do not use the test leads if they are cracked or 
have exposed metal. 

 

Do not exceed the test lead’s voltage and 
current rating. 

 

Make sure test leads are fully mated to the 
instrument and accessories. 

 

Keep fingers behind the finger guards 

 

Adhere to local and national safety codes.  

 

Individual protective equipment must be used 
to prevent shock and injury. 

International Symbols 

Symbol Meaning 

Earth ground 

Hazardous voltage 

Risk of danger. Important information. See 
instruction sheet. 

Conforms to European Union directives 

Double Insulated 

 