Fluke ScopeMeter 199C Скачать руководство пользователя страница 5

 

5.

 

Press F4 save, and your setup and the screen that was on the scope 
will be saved in the location you selected.   

C.

 

Recalling a Setup. 
1.

 

Press the round “Save/Print” button to open the save menu. 

2.

 

Press the F2 Recall button. 

3.

 

Use the up/down arrows to highlight the location (out of 15 choices) 
where you saved the setup. 

4.

 

Press F2 Recall Setup button to recall you saved setup. 

5.

 

You will be automatically be brought to a screen w/ your setup. 

D.

 

To capture a series of fencer transients.   
1.

 

Attach you 1:1 probes to the Input A connector. 

2.

 

Turn on the scope.   

3.

 

Connect the probe between the two points of the selected cow contact 
location, preferable at close to (in) the location and w/o other wires 
attached to the location. 

4.

 

You will have to adjust the trigger level, voltage scale, and time scale 
to capture what you want.  

a.

 

Start with the voltage scale set to 1 Volt per division, the time scale 
set at 1 second per division, and the trigger set around a couple of 
volts.  (Use the 1 second per division scale because the fencers 
generally put out 1 pulse per second, making it easy to identify that 
what you are capturing is a fencer or trainer transient.) 

b.

 

If nothing shows up, open the trigger menu by pressing the 
rectangular Trigger button. 
1)

 

Use the up/down arrows to adjust the trigger level until you 
get something.   

2)

 

If you still get nothing, change the voltage scale by pushing the 
Input A Range button to change the scale to higher or lower 
values.  Note the scale per division changes in the red box on 
the lower left side of the screen. 

c.

 

If the screen is overloaded, with signals appearing off screen, 
change the input scale by pushing the Range button to raise the 
scale.  Raise it until the entire signal appears on the screen.   

5.

 

Press the Clear Menu button to clear the trigger menu, once you are 
satisfied w/ your snapshot. 

6.

 

To save the screen, press the Save/Print button, press F1 to open the 
save menu, use the up/down arrows to highlight the location you 
wish to save the screen, and press save. Press the Clear Menu button 
to clear the Save/Print menu. You are now ready to go again.   

Содержание ScopeMeter 199C

Страница 1: ...creens This is the case whether triggering for the scope is set to automatic trigger one shot or on trigger Each refresh will result in a screen stored in the memory B To look at what has been saved press the Replay button 1 A control panel will appear at the bottom of the screen allowing you to scroll through each screen one at a time for a close look scroll back one at a time or hit play and rap...

Страница 2: ... the lower portion of the display G Press the F2 button below the Recall designation or the F4 button below the View designation Downloading Saved Screens to a File or Document A Connect the scope to your computer via the optical coupler supplied with the scope meter B Open Fluke View The softwear will automatically prompt you about your desire to connect to the Instrument C Turn on the Scope Mete...

Страница 3: ...are A button to turn on the channel A menu A blue menu will appear at the bottom of the screen b Alternately pushing the F1 button will turn input A on and off Turn input A on c Alternately pushing the F2 button will alternate the coupling between DC direct coupling and AC Turn on DC d Generally we will be using the 1 1 probe Pushing the F3 button will open up the probe menu 1 From the open probe ...

Страница 4: ...n Edge submenu Use the up down arrows to highlight single shot and press F4 enter to select single shot 3 You will be automatically moved over to Noise Rejection Filter submenu Use the up down arrows to highlight OFF press F4 Enter to select and you will automatically be brought back to the main trigger menu f Press the Clear Menu button to clear the trigger menu 7 Input to Channel B Input a Press...

Страница 5: ...around a couple of volts Use the 1 second per division scale because the fencers generally put out 1 pulse per second making it easy to identify that what you are capturing is a fencer or trainer transient b If nothing shows up open the trigger menu by pressing the rectangular Trigger button 1 Use the up down arrows to adjust the trigger level until you get something 2 If you still get nothing cha...

Страница 6: ...you can change the time scale all the way down to 5 nanoseconds division ns and still have the glitch detect activated 8 Once you capture a signal you are happy w by alternately pressing the hold run button and changing the time scale and voltage scale as necessary you can save the screen as previously described Press the Save Print button press F1 to open the save menu use the up down arrows to h...

Страница 7: ...w softwear c Clicking on the camera icon will bring in the active scopemeter screen to the Flukeview softwear Clicking on the camera w the question mark will let you bring in screens from the memory locations d Once in Flukeview you may 1 Save the screens as bitmap or fvf files at a location of your choosing FVF files are for Flukeview and can only be viewed by the softwear A BMP is better 2 You c...

Отзывы: