ScopeMeter® Test Tool
Calibration Adjustment
49
4. Set the 5502A to operate (OPR).
5. Press
3
to start the calibration.
Wait until the display shows the calibration status
:READY
.
6. Press
2
to select the next calibration step, set the 5502A to the next calibration
point, and start the calibration.
7. Continue through all calibration points for
Multimeter Ohm Gain
until the display shows the
status
CL 0916: READY
.
When you are finished, set the 5502A to Standby.
Continue at the
Save Calibration Data and Exit
section.
Save Calibration Data and Exit
To save the calibration data and exit the Maintenance mode:
1. Remove all test leads from the Test Tool inputs.
2. Press
4
(SAVE). The Test Tool shows on the display:
Calibration data valid.
Save data and exit maintenance mode?
Note
Calibration data valid indicates that the calibration adjustment procedure is performed
correctly. It does not necessarily mean that the Test Tool meets the specifications.
3. Press
4
(YES) to save and exit.
Note
After saving the calibration data, the calibration number and date updates if the
calibration data changes and the data are valid. The calibration number and date do not
change if:
the calibration mode is entered and left without doing a calibration adjustment
only the probe calibration was done
you press
3
(NO), the Test Tool returns to the calibration mode
you can either calibrate the Test Tool again, or press
4
(EXIT), (YES) to save and
exit
Error messages:
WARNING: Calibration data not valid.
Save data and exit maintenance mode?