ScopeMeter® Test Tool
Performance Verification
25
b. Press
P
.
c. Press
2
(– READING …) and select with
1
(– READINGS) and
1
(READINGS) on
A | V ac
.
d. Press
J
to select autoranging (AUTO in upper right LCD edge).
e. Use
j
and
k
to change the sensitivity range to select manual sensitivity ranging, and
lock the Input A sensitivity range on
500 mV/div
. AUTO in upper right LCD edge
becomes ½ AUTO.
f. Use
l
to move the Input A trace zero to the center grid line.
3. Set the 5502A to source a sine wave and to the first test point in
.
4. Observe the Input A reading and check that it is within the range shown under the
appropriate column.
5. Continue through the test points.
When you are finished, set the 5502A to Standby.
Input B AC Voltage Accuracy (HF) and Bandwidth Test
To test the Input B high frequency automatic scope ac voltage measurement accuracy and the
bandwidth:
1. Connect the Test Tool to the 5502A. See
.
2. Select the Test Tool setup:
). Press
S
,
2
(RECALL) and
select
SETUP
, press
E
, select the setup name, and press
E
to recall the
setup.
b. Press
P
.
Table 10. HF AC Voltage Verification Points Input A
UUT Model
5502A SC... MODE levsine
UUT
Reading A
Voltage
Frequency
All
2.545 Vpp
1 MHz
835 mV to 965 mV
All
2.545 Vpp
25 MHz
790 mV to 1.010 V
Models 60 MHz
2.545 Vpp
60 MHz
>630 mV
Models 100 MHz
2.545 Vpp
100 MHz
>630 mV
Models 200 MHz
2.545 Vpp
200 MHz
>630 mV
Models 500 MHz
2.545 Vpp
500 MHz
>630 mV