ScopeMeter® Test Tool
Performance Verification
21
4. Use
V
to select the time base in
5. Set the 5502A to source the leveled sine wave for the appropriate Test Tool model.
6. Adjust the 5502A output voltage until the displayed trace has the trigger amplitude
indicated under the last column of
.
7. Verify that the signal is well triggered.
If not, press
T
and use
3
to enable
h
i
for manual Trigger Level
adjustment. Adjust the trigger level and verify that the signal is triggered. The
trigger icon (
) indicates the trigger level.
8. Continue through the test points.
When you are finished, set the 5502A to Standby.
Table 9. Input Trigger Sensitivity Test Points
UUT Model
UUT Time base
5502A SC... MODE levsine
UUT Trigger
Amplitude
Initial Input Voltage Frequency
All
200 ns/div
100 mV pp
5 MHz
0.5 div
190-062
10 ns/div
400 mV pp
60 MHz
1 div
10 ns/div
800 mV pp
100 MHz
2 div
190-102/190-104
10 ns/div
400 mV pp
100 MHz
1 div
10 ns/div
800 mV pp
150 MHz
2 div
190-202/190M-2/
190-204/190M-4
10 ns/div
400 mV pp
200 MHz
1 div
10 ns/div
800 mV pp
250 MHz
2 div
190-502/190-504/
MDA-550-III
2 ns/div
400 mV pp
500 MHz
1 div
2 ns/div
800 mV pp
600 MHz
2 div