Input A, B, C, D DC Voltage Accuracy Test
XW
Warning
To prevent possible electrical shock, fire, or personal injury, ensure that the
calibrator is in standby mode before making any connection between the
calibrator and the Test Tool. Dangerous voltages are present on the calibration
source and connection cables during these steps.
Note
The test steps for channel C and D are only for the 4-channel models.
To verify the automatic dc voltage scope measurement:
1. Connect the Test Tool to the 5502A as shown in see
Figure 3. Test Tool Inputs A, B, C, D to 5502A Normal Output
2. Select the Test Tool setup:
). Press
S
,
2
(RECALL) and
select
SETUP
.
b. Press
E
, select the setup name, and press
E
to recall the setup.
c. Press
A
, press
4
(INPUT A OPTIONS...), and select
Attenuator: Normal |
Bandwidth: 10 kHz
for the available setting in the ScopeMeter.
d. Press
B
, press
4
(INPUT B OPTIONS...), and select
Attenuator: Normal |
Bandwidth: 10 kHz
for the available setting in the ScopeMeter.
e. Press
C
, press
4
(INPUT C OPTIONS...), and select
Attenuator: Normal |
Bandwidth: 10 kHz
for the available setting in the ScopeMeter.
f. Press
D
, press
4
(INPUT D OPTIONS...), and select
Attenuator: Normal |
Bandwidth: 10 kHz
for the available setting in the ScopeMeter.
g. Press
I
to clear the softkey menu and the full 8-divisions display.
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
PM9091
PM9081
CONNECT TO CHANNEL A, B, C, D IN PARALLEL
50
:
TERMINATION
NO
A
D
C
B
PM9093
PM9092
A
B
METER
P
M
9
0
9
3
PM9092
2 CHAN. + METER
4 CHANNELS