B-1
Appendix B
Summary of Error Messages
Summary of Error Messages
Negative error numbers are from the Standard Commands for Programmable Instruments
(Version 1991.0).
Table B-1. Summary of Error Messages
Value
Description and Corrective Action
0 No
Error.
-103
Invalid Separator. Check punctuation in the SCPI command.
-104
Data Type. The type of parameter data is incorrect. Check for numeric versus string
data.
-109
Missing Parameter. No valid parameter was found for the SCPI command.
-110
Command Header. The command name is not valid.
-113
Command Unknown. The command specified does not exist.
-114
Header Suffix. The numeric suffix for the command name is out of range.
-201
Remote Required. Remote command conflicted with local interface. Return local
interface to main Menu.
-221
Settings Conflict. The command could not be executed due to the current state of
the Calibrator. Some commands cannot be executed while a program, self-test, or
calibration is in progress.
-222
Out of Range. The value is not within the valid range. For pressures, check your
high and low limits.
-281
Cannot create program. Program memory is full.
-282
Illegal Program Name. The name specified is not valid or does not exist.
-284
Program Currently Running. The command cannot be executed while a program is
running.
-285
Program Syntax Error. The syntax of the program definition is not correct.
Содержание RUSKA 7350
Страница 8: ...RUSKA 7350 Users Manual vi ...
Страница 10: ...RUSKA 7350 Users Manual viii ...
Страница 16: ...RUSKA 7350 Users Manual 1 6 ...
Страница 22: ...RUSKA 7350 Users Manual 2 6 ...
Страница 52: ...RUSKA 7350 Users Manual 4 28 ...
Страница 62: ...RUSKA 7350 Users Manual 5 10 ...
Страница 72: ...7350 Users Manual 7 4 ...
Страница 76: ...RUSKA 7350 Users Manual A 4 ...