RSE300, RSE600
Users Manual
4
Operation
The Imager and SmartView R&D need a stable network connection to work properly. Set up
your network without interference from other systems. Do not connect the Imager to a Virtual
Private Network (VPN). SmartView R&D is a data-intensive application. Use of other data-
intensive applications (streaming audio or video) on the PC or the same network system can
cause loss of data.
Note
All thermal imagers need sufficient warm-up time for accurate temperature
measurements and best image quality. Fluke recommends a warm-up period of
60 minutes for optimal performance.
To set up the Product:
1. Attach the antenna.
2. Stabilize the Product either on a flat surface or a tripod.
3. Connect the M12 end of the ethernet cable to the M12 ethernet jack on the Product. See
.
a. Fully insert the ethernet cable into the ethernet jack.
b. Rotate the ring 180 º clockwise. Connect the M12 connector end of the ethernet cable
to the Product.
Figure 1. M12 Ethernet Cable Connection