PV350 English Instruction Sheet 

Page 

 

To ensure that the module is used safely, follow all

safety and operating instructions in this Instruction 

Sheet. If the module is not used as described in this 

Instruction Sheet, the safety features of the module 

might be impaired.

Do not use the module if the module or fittings look

damaged, or if you suspect that the module is not

operating properly. 

 This symbol on the module indicates that an

operator should refer to an explanation in the

instruction sheet.

Follow accepted safety procedures for the systems on

which pressure or vacuum measurements are being

made.

Materials Compatibility 

The PV350 is compatible with industrial gases and liquids that are 

compatible with 316 stainless steel. To determine the compatibility of a 

liquid or gas, refer to the manufacturer's specifications. When making 

measurements on more than one type of pressure system, make sure the 

fluids or gases in those systems are compatible. If they are not, 

contamination or damage to the system being measured may result.  

Cleaning 

Clean the module using a soft cloth and mild soap and water. Do not 

immerse the PV350 in liquids. 

 

Clean the transducer and connectors after each use with a cleaner 

appropriate for the liquid or gas connected during measurement. Refer to 

the manufacturer’s instructions for the proper cleaning material. 

 

Measuring Pressure and Vacuum 

Caution 

The sensor diaphragm can be damaged by solid or sharp 

objects. Never insert any object into the inlet port. 
Clean the sensor immediately after use with appropriate 

solvents (refer to the  solvent manufacturer's instructions). 

Note 

Fluke recommends that Teflon sealing tape (or its equivalent) 

be used in fitting threads to ensure good seals between the 

PV350 and any hoses or fittings. 

1.

Plug the PV350 into the DMM (red polarity dot to volts input) and set

the multimeter to mV dc.

2.

Check the battery voltage (see Testing the Battery).

3.

Press the English/Metric button to select the desired units.

4.

Slide the switch to the desired setting.

5.

Before applying pressure, zero the module by turning the ZERO

potentiometer until the DMM displays zero. (Zero on the range you

will be using and rezero the DMM whenever the scale is changed.)

Содержание PV350

Страница 1: ...ith min max function become a datalogger capable of measuring peak high and peak low Box Contents Pressure Vacuum Transducer Module 1 4 in male flare to 1 4 in female pipe thread connector 9 V alkaline battery Instruction Sheet Quick Reference Guide Safety Information Warning To avoid eye injury always wear approved safety glasses when working with this instrument To avoid injury from pressure exp...

Страница 2: ...sured may result Cleaning Clean the module using a soft cloth and mild soap and water Do not immerse the PV350 in liquids Clean the transducer and connectors after each use with a cleaner appropriate for the liquid or gas connected during measurement Refer to the manufacturer s instructions for the proper cleaning material Measuring Pressure and Vacuum Caution The sensor diaphragm can be damaged b...

Страница 3: ...le s specified accuracy 6 Apply pressure vacuum to the PV350 and read the DMM display When measuring pressure if the DMM displays OL switch the multimeter range from millivolts to dc volts Increasing the range to dc volts moves the decimal position three places to the left 500 psi displays as 0 500 V When measuring vacuum the DMM displays a minus sign Testing the Battery 1 Turn the DMM and the mod...

Страница 4: ...RANGE HOLD H mA A mV V V OFF A mA A 10A MAX FUSED ZERO SMOOTH Hz RPM DUTY mA A RPM ALERT TRIGGER _ ms PULSE COM 1000V MAX 400mA MAX FUSED DC mV 4000 mV 20 2 4 6 3 8 2 4 6 8 4 V RPM METRIC ENGLISH Z PV500 PRESSURE VACUUM MODULE cmHg inHg kPa psi OFF Figure 3 Pressure for HVAC R Super Heat Measurements 88 AUTOMOTIVE TESTER MIN MAX RANGE HOLD H mA A mV V V OFF A mA A 10A MAX FUSED ZERO SMOOTH Hz RPM ...

Страница 5: ...a cm Hg or in Hg Resolution Determined by the resolution of the measuring device RFI Specification Typically 1 error from 80 to 200 MHz at 3 V m Electromagnetic Compatibility EMC International IEC 61326 1 Portable IEC 61326 2 2 CISPR 11 Group 1 Class A Group 1 Equipment has intentionally generated and or uses conductively coupled radio frequency energy that is necessary for the internal function o...

Страница 6: ...Standard 9 V Alkaline battery NEDA 1604 IEC 6LR61 Battery Life 300 hours typical Electrical Cable Black PVC will withstand up to 105 C Cable Length 8 ft Pressure Port Connection 1 4 in NPT Connector Brass 1 4 in male flare to 1 4 in female pipe thread List of Replaceable Parts Part Number Description 614487 2813197 2816172 913207 926873 926881 Battery Primary 9 V 0 15 mA Case Bottom PV350 Case Top...

Отзывы: