PG9602™ OPERATION AND MAINTENANCE MANUAL
© 2011 Fluke Calibration
Page
10
2.1.1.2
Piston-Cylinder Assemblies
The piston-cylinder modules are shipped in Acetal bullet cases that are packed in
corrugated containers with custom foam inserts.
Open the corrugated containers and remove the piston-cylinder modules
and accessories.
The bullet cases screw open by turning the lid counterclockwise.
2.1.1.3
Vacuum Reference Hardware (optional)
The optional PG9000 Vacuum Reference Hardware includes the vacuum Bell
Jar, reference vacuum sensor with power supply and interconnect cables and
fittings necessary for vacuum operation.
The PG9000 Bell Jar is packed in a large wooden crate (66 cm x 66cm x 71 cm).
The vacuum reference sensor, power supply and associated fittings and cables
are packed in a reusable molded Vacuum Measurement Kit transit case.
Remaining fittings and adaptors are shipped in the PG9000 accessories
corrugated container (see Section 2.1.2.3).
2.1.1.4
AMH-100, Automated Mass Handler (optional)
AMH-100-VAC is delivered in a corrugated container with foam cut outs to hold
the AMH in place. Accessories are bagged together and packaged in a separate
corrugated container with other PG9000 system accessories.
Remove all parts from the shipping boxes. See Section 2.1.2.4 for more detail.
2.1.1.5
AMH Mass Set (optional)
PG9000 AMH masses are shipped in reusable, molded shipping and storage
cases. One of the cases contains the binary masses, binary mass carrier, mass
bell and lifting shaft/trim mass tray. The other cases contain the main masses of
10 kg (see Section 2.1.2.2 for mass set content detail). Each mass is packed in a
plastic bag and then placed in a protective shipping insert. The PG9000 AMH
masses should be removed from their shipping cases and inventoried when
setting up the PG9000 system (see Section 2.1.1.6 for mass set content detail).
Caution
The stability over time of PG9000 pressure measurements is a
function of the stability of the masses loaded on the cylinder.
Precautions should be taken in handling the masses to minimize
influences that may change their mass. This includes always
wearing protective gloves when handling the masses to avoid
contaminating them with body oils and perspiration. Protective
gloves are provided in the accessory kits of PG9000 Platforms.
2.1.1.6
Manual Mass Set
The PG9000 manual masses are shipped in reusable, molded shipping and
storage cases. The PG9000 masses should be removed from their shipping
cases and inventoried when setting up the PG9000 system (see Section 2.1.2.2
for mass set content detail). The default manual mass set is the type MS-7002
which has a 4 kg makeup mass.
Содержание PG9000 Series
Страница 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Страница 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Страница 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Страница 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Страница 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...