Hydraulic Comparison Test Pump 

 

Maintenance and Servicing 

4.

 

The white, anti-extrusion ring (21) is a PTFE spiral, and can be removed by 

“unwinding” it from the rambler. 

5.

 

When removing the rambler seal (20), take care not to use any tool that may have a 

sharp edge that will scratch the surfaces of the rambler, otherwise it may leak when 

reassembled. 

6.

 

The replacement rambler seal can be eased over the front of the rambler, and into 

the groove. 

7.

 

Similarly, the new anti-extrusion ring can be “wound” into the groove in the rambler, 

behind the rambler seal. 

8.

 

If the rambler has been separated from the lead screw, ensure that the ball (23) is 

correctly fitted before reassembly. 

9.

 

Ensure that the rambler assembly is correctly located on the end of the lead screw 

assembly. Carefully introduce the rambler into the open end of the barrel; making 

sure that it does not tilt when entering the barrel. 

10.

 

Push the lead screw assembly fully in to the barrel, ensuring that the key in the nut 

locates correctly in the slot in the barrel. 

11.

 

Re-tighten the barrel union nut. 

W

 Caution 

If the lead screw assembly (24) shows signs of excessive wear, 

then it is very likely that the associated components have worn 

also; therefore, the screw press assembly is available as a 

spare part. 

Complete Disassembly 

1.

 

Remove the screw press assembly as described above. 

2.

 

Unscrew and remove valve stem (13), taking care not to lose spring (12) and nylon 

washer (11). 

3.

 

Remove reservoir cover (10). 

4.

 

Unscrew locknut (9), and remove reservoir (8). 

5.

 

Unscrew valve body (6), taking care not to lose bonded seal (7) or O-ring (5). 

6.

 

Remove banjo bolts (1), together with banjo test ports (3), taking care not to lose 

O-rings (2). 

7.

 

Disconnect the stand (15) from the bench. Remove screws (14), and tilt 

manifold/barrel assembly downward and pass it out of the bottom of the stand.     

8.

 

To remove the barrel (19), the locknut (18) must be loosened approximately 1/2 turn.  

The barrel can then be unscrewed from the manifold (16). 

9.

 

Before re-fitting the barrel, ensure that the barrel seal (17) is correctly located in the 

counter-bore in the front of the barrel. Screw the barrel fully in to the test station, and 

secure with the locknut. 

10.

 

Inspect all seals for wear or damage — replace as necessary. 

Содержание P5500 Series

Страница 1: ...uke Corporation All rights reserved Printed in USA Specifications are subject to change without notice All product names are trademarks of their respective companies P5514 P5514 V P5514 EP Hydraulic C...

Страница 2: ...vice center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destination Fluke assumes no...

Страница 3: ...ressed Liquid 2 Personal Protective Equipment 2 Symbols Used in this Manual 2 Connections 3 Connect the Gauge 4 Connect the Test Port Insert 4 Operation 5 Priming 5 Operation 6 Maintenance and Servici...

Страница 4: ...P5514 P5514 V P5514 EP Users Manual ii...

Страница 5: ...iii List of Tables Table Title Page 1 Operating Fluid Compatibility 2 2 Symbols 3 3 Test Port Insert Parts List 5 4 Seal Kit Part Numbers 6 5 Screw Press Assembly Parts List 10...

Страница 6: ...P5514 P5514 V P5514 EP Users Manual iv...

Страница 7: ...v List of Figures Figure Title Page 1 Connecting the Gauge 4 2 Test Port Insert 5 3 Reservoir and Test Port Assemblies 9 4 Screw Press Assembly 9 5 Angle Adapter 11 6 Pointer Remover Punch 11...

Страница 8: ...P5514 P5514 V P5514 EP Users Manual vi...

Страница 9: ...acy is maintained The comparison test pump should be mounted securely to a stable workbench or similar surface Four mounting holes are provided in the pump stand for this purpose How to Contact Fluke...

Страница 10: ...ns that are not related to any specific procedures and do not appear elsewhere in this publication These are recommended precautions that personnel must understand and apply during equipment operation...

Страница 11: ...fluids and protect the system from particulates we recommend the use of a Liquid to Liquid Separator see Section 5 Ancillary Equipment W Warning DO NOT use Teflon PTFE tape on these connections as th...

Страница 12: ...the position to face forward hold the gauge adapter and turn the instrument COUNTER CLOCKWISE so that it faces forward 4 Hold the instrument steady while turning the gauge adapter COUNTER CLOCKWISE un...

Страница 13: ...Open reservoir valve 13 one turn counter clockwise and turn screw press fully in 2 Close valve 13 and turn screw press fully out 3 Open valve and turn screw press fully in Note During this operation b...

Страница 14: ...ution Never release the system pressure by using the reservoir valve alone as sudden depressurization will shock the system which may cause damage to sensitive instruments attached to it Maintenance a...

Страница 15: ...t in the barrel 11 Re tighten the barrel union nut W Caution If the lead screw assembly 24 shows signs of excessive wear then it is very likely that the associated components have worn also therefore...

Страница 16: ...n the barrel is worn also If the bore is worn or scratched it will not seal correctly and may leak under pressure 11 Re assembly is the reverse of the above Ensure that all seals and sealing surfaces...

Страница 17: ...Hydraulic Comparison Test Pump Maintenance and Servicing 9 20 22 23 24 25 25 26 26 21 16 17 19 18 gjq004 eps Figure 4 Screw Press Assembly...

Страница 18: ...lve Body 3923785 7 Bonded Seal 3862349 3922096 3922043 8 Reservoir 3919463 9 Locknut 3918244 10 Reservoir Cover 3918350 11 Washer 3916458 12 Spring 3922786 13 Valve Stem Assembly 3885982 14 Screw 3909...

Страница 19: ...ould be used The angle adapter fits directly onto the test station converting it through 90 degrees allowing the standard adapters to be used gjn022 bmp Figure 5 Angle Adapter Pointer Remover Punch P5...

Страница 20: ...P5514 P5514 V P5514 EP Users Manual 12...

Отзывы: