2. Push and hold the

button while powering on the Product to 

enter compass calibration mode. The Product shows CAL and the 

compass icon.

2 m (6 ft)

1. Do the calibration outdoors, at least 6 feet (2 m) away from any 

metal objects.

2

5. Turn off the Product. The Product stores the calibration data.

Supplement

4. Slowly turn the Product around on the vertical axis (approximately 

2 seconds per movement). Continue turning the product while 

slowly rotating your body 360 degrees for 1 complete rotation at a 

steady pace, take at least 1 minute to complete the body rotation.

360°

3. Slowly turn the Product around all three axis in the space using 

figure 8 movement (approximately 3 seconds per movement). 

Continue turning the product while slowly rotating your body 360 

degrees for 3 complete body rotations at 

steady pace, take at 

least 1 minute to complete the body rotation. 

360°

Change #1,

Add a new section, 

Compass Calibration

 before the 

Maintenance

 section. 

Compass Calibration

The Product contains a compass. The compass is affected by metal in the 

environment. Recalibrate the compass:

- Before the first use.

- Each time you change the batteries.

- When you change locations.

- When there is a significant change in the environment. (For example new 

metal fittings.) 