IDA-5
Users Manual
22
Channel 1 PCA/Dual Flow Graph
80
Ave
ml/h
Bolus Count
3
Basal Avg
0
Bolus Avg
65.00
Total Vol
6.00
ml:
15 min
VIEW DETAIL
Press ESC for Status Screen
gir20.eps
Figure 9. PCA Graph Screen
Caution
To avoid inaccurate readings, always repeat a test when
“Bubble” or “Air Lock” is shown on the display while a test is
running. See the “Troubleshooting” section of this manual.
Dual Flow Pump Test
During a test, a dual flow pump can be thought of as a PCA pump that supplies a large,
single bolus.
The primary flow rate must be recorded in the basal flow rate field of the PCA
Information screen. Record the secondary volume in the bolus volume field.
Templates
The Product helps automate and standardize the testing of infusion devices. Templates
are used by the Product to control the sequence of tests done on an applicable type of
infusion device.
Before you can do this, a template for the type of infusion device you will test must be
defined. It is necessary to define a template only once for a particular type of infusion
device. Refer to the device’s manufacturer or service documentation for necessary test
parameters.
The subsequent section shows an example of a generic syringe pump that requires you
test at a flow rate of 20 ml/h for a volume of 10 ml over 30 minutes, and 50 ml/h for
10 ml, both with a tolerance of 5 %. The occlusion must be tested at 50 ml/h with an
alarm pressure of 12 psi.