
•
De-energize the installation on which current will be measured
or adopt safe operating procedures
during application and
removal of the current probe.
•
Use extreme caution when working around bare
conductors or
bus bars.
•
Do not use the Current Probe to measure bare conductors
carrying a voltage from 30 V up to 1000 V unless you are
wearing protective clothing suitable for high-voltage work.
Contact with the conductor could result in electric shock. Always
use appropriate
equipment for personal protection.
•
Use caution when working with voltages above 60 V dc, 30 V ac
rms or 42 V ac peak. Such voltages pose a shock hazard.
Symbols
Do not apply around or remove from
HAZARDOUS
LIVE
conductors.
Product is protected by double insulation.
Risk of Danger. Important information. See
Instruction Sheet.
Risk of Electric Shock.
Conforms to relevant European standards.
Specifications
SAFETY
i430flex
(INPUT and
OUTPUT)
Complies with American industry
standards UL61010B-1 &
UL61010B-2-032 & European
standards EN/IEC 61010-1 2nd
Edition & EN/IEC 61010-02-032.
ELECTRICAL SPECIFICATIONS
Measuring range
30 to 3000 A ac
Maximum non-destructive
current
100 kA
Output signal
85 mV at 1000 A rms, 50 Hz
Basic accuracy
±1% of reading at 25
°
C, 50 Hz
Linearity
±
0.2% of reading at 10%...100% of
range
Noise
< 1 mV rms at 10 Hz…7 kHz
Additional errors:
•
with temperature
(0 to +70 °C,
32 to 158 °F)
0.08% of reading /°C,
0.144% of reading /°F
•
with position of conductor
in the probe window
±2% of reading (bus
≥
2.5 cm or
1 inch from coupling)
Phase shift 45 to 65 Hz
< ±1°
Bandwidth (-3dB)
10 Hz to 7 kHz
GENERAL SPECIFICATIONS
Weight 250
g
Transducer length
610 mm (24 in)
Transducer diameter
12 mm (0.49 in)
Minimum bending radius
40 mm (1.57 in)
Cable length from
transducer to BNC
2.5 m (98.4 in)
Temperature, operating
-20 to +90°C (-4 to +194°F)
Temperature, non-operating -40 to + 105°C (-40 to +221°F)
Relative Humidity, operating 15 to 85%, non condensing
Altitude, operating
3000 m
Derate to 1000 V CAT II / 600 V
CAT III / 300 V CAT IV above
2000 m
Altitude, storage
12 km
Instrument Compatibility
The i430flex can only be used with the Fluke 430 series Power
Quality Analyzers.
Using the i430flex
To use the i430flex, follow these instructions:
1.
Connect the BNC connector of the i430flex to the desired input
on the Fluke 430 series Power Quality Analyzer.
Figure 1. Connecting the flexible probe around the conductor.
2.
Connect the flexible probe around the conductor (See
Figure 1.)
3.
Center the conductor perpendicularly inside the flexible probe
area. If this is not possible, an additional measurement error of
±2% of reading can occur.
4.
Avoid taking measurements close to other current-carrying
conductors if possible.
5.
Make sure that the arrow marked on the probe coupling points
toward the correct orientation for correct phase display on the
instrument.
6.
Keep the probe coupling more than 2.5 cm (1 inch) away from
the conductor.
7.
Observe the current value and waveform on the instrument’s
display.
Warning
If the selected current probe of the Fluke 430 series
Power Quality Analyzer is not the i430flex, the
Fluke 430 will display false and misleading readings.
Maintenance
Before each use, assure continued safety by inspecting the flexible
probe and its latching system for any damage. Pay particular
attention to the insulation surrounding the flexible probe. An i430flex
under warranty will be promptly repaired or replaced (at Fluke's
discretion) and returned at no charge.
Cleaning and Storage
•
The flexible probe and its latching system require no special
care. Ensure that no foreign body obstructs the latching
mechanism.
•
Clean the i430flex with a damp cloth and a mild detergent. Do
not use abrasives, solvents, or alcohol.
If your i430flex does not work
If the i430flex does not perform properly, use the following steps to
help isolate the problem:
•
Inspect the coupling system for any damage. If any foreign
material is present, the coupling system will not close properly
and errors will result.
•
Inspect the cable between the flexible probe and the Power
Quality Analyzer for any damage.
•
Check if the i430flex is the selected current probe on the
Power Quality Analyzer (Clamp i430flex).
•
Verify that the function and range selection on the Power
Quality Analyzer is correct.