Operation
To use:
1. Connect the clamp cable to the current inputs on the measurement
instrument. The instrument automatically recognizes the clamp and
selects the default range of the instrument for the probe.
2. Make sure the measurement instrument is set to the correct range.
3.
Connect the jaws around the conductor to be measured.
Warning
To prevent electrical shock, fire, or personal injury, hold the
Product behind the tactile barrier. See Table 1.
Table 1. i400s-EL Setup
A
D
B
C
E
F
Item
Description
A
Single Current Carrying Conductor
B
Load Direction Arrow
C
Alignment Marks
D
Tactile Barrier
E
2 m (6.6 ft) Output Cable
F
Connector
Maintenance
Warning
To prevent electrical shock, fire, or personal injury:
• Before each use, examine the Product. Look for cracks or
missing pieces of the clamp housing or output cable insulation.
Also look for loose or weakened components. Carefully
examine the insulation around the jaws.
• Do not use the Product if it is altered or damaged.
If the Current Clamp does not work or perform properly:
1.
Inspect the jaw mating surface for cleanliness. If any foreign material
is present, the jaw will not close properly and measurement errors
will result.
2. Verify that the function selection and range on the measurement
instrument are correct and adjusted to the sensitivity of the Current
Clamp.
Cleaning
Periodically wipe the case with a damp cloth and mild detergent.
Caution
To avoid damaging the Current Clamp, do not use abrasives
or solvents to clean the clamp.
Open the jaws and wipe the magnetic pole pieces with a lightly oiled
cloth. Do not allow rust or corrosion to form on the magnetic core ends.
Specifications
Measuring Range ........................... 400 mA ac to 40 A ac / 4 A ac to 400 A ac
Crest Factor ....................................
≤3
Accuracy at Reference
Condition
[1]
...................................... ±0.5% of reading
[1] Reference Condition:
• 400 A 50/60 Hz
• Environmental: 23 °C ±5 °C, no external electrical/magnetic field, RH 65 %
• Primary conductor in center position
Accuracy 174x + Clamp
400 A range ................................ ±(0.7 % of r 0.1 % of range)
40 A range .................................. ±(2 % of r 0.2 % of range)
Phase Shift
<400 mA .....................................
unspecified
400 mA to 1 A ............................. < ±3.5°
1 A to 2 A .................................... < ±2.5°
2 A to 10 A .................................. < ±1.5°
10 A to 400 A .............................. < ±1°
Temperature Coefficient over
Operating Temperature Range ....... 0.01 % of reading / °C
0.0056 % of reading / °F
Influence of Adjacent Conductor
.....
≤9 mA/A (@ 50/60 Hz)
Influence of Conductor Position
in Jaw Opening ............................... ±1.0 % of r 0.05 A
Bandwidth ....................................... 10 Hz to 2.5 kHz
Safety
General ...................................... IEC 61010-1: Pollution Degree 2
Measurement ............................. IEC 61010-2: CAT III 1000 V/ CAT IV 600 V
Size (H x W x L) .............................. 150 mm x 70 mm x 30 mm
(6 in x 2.8 in x 1.2 in)
Maximum Conductor Size .............. 32 mm (1.2 in)
Output Cable Length ...................... 2 m (6.6 ft)
Weight ............................................ 272 g (9.6 oz)
Temperature
Operating ................................... 0 °C to 50 °C (32 °F to 122 °F)
Non-operating ............................ -20 °C to +60 °C (-4 °F to +140 °F)
Relative Humidity, operating ........... 10 °C to 30 °C: 95 %
30 °C to 40 °C: 75 %
40 °C to 50 °C: 45 %
Operating Altitude ........................... 2000 m (6500 ft) up to 4000 m (13 000 ft) derate
to CAT II 1000 V / CAT III 600 V / CAT IV 300 V
Storage Altitude .............................. 12 km (40 000 ft)
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for one
year from the date of purchase. This warranty does not cover fuses, disposable
batteries, or damage from accident, neglect, misuse, alteration, contamination,
or abnormal conditions of operation or handling. Resellers are not authorized to
extend any other warranty on Fluke’s behalf. To obtain service during the warranty
period, contact your nearest Fluke authorized service center to obtain return
authorization information, then send the product to that Service Center with a
description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH
AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED.
FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR
CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR
THEORY. Since some states or countries do not allow the exclusion or limitation
of an implied warranty or of incidental or consequential damages, this limitation of
liability may not apply to you.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Fluke Europe B.V
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
ООО «Флюк СИАЙЭС»
125167, г. Москва,
Ленинградский проспект дом 37,
корпус 9, подъезд 4, 1 этаж
11/99