i3000/i2000 English Instruction Sheet
Page 8
Measuring Current
XW
Warning
•
To avoid electric shock or injury, read Safety
Instructions before operating this product
•
Ensure that the conductor to be tested is de-
energized
To measure current:
1.
Connect the output of the electronics to the input of an
oscilloscope or other data recording device.
XW
Warning
To avoid electric shock or injury, the flexible current
probe is not for use on conductors with a potential
of over 600 V.
2.
Wrap the flexible measuring head around the conductor
to be tested in a close coupling manner.
3.
Energize the circuit under test.
4.
For most accurate measurement, centre the flexible
head around the conductor.
5.
Locate coupling away from the nearby conductors.
WX
Warning
To avoid electric shock or injury:
•
Do not use the flexible current probe to
measure bare conductors unless wearing
protective clothing suitable for high voltage
work.
•
Always use appropriate equipment for personal
protection. When installed on bare
conductors/busbars, the product must be within
a suitable enclosure.