Manual Supplement
i1000s Calibration
2
7/01
0
•
1
2
3
4
5
6
7
8
9
ENTER
M
k
m
V
Hz
FIELD
EDIT
/
+
F
OPR
STBY
EARTH
SCOPE
EXGRD
MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
µ
n
p
W
dBm
sec
¡C
POWER
A
MULT
x
DIV
÷
MODES
MORE
5520A CALIBRATOR
HI
LO
TRIG
GUARD
TC
20A
NORMAL
AUX
SCOPE
OUT
V, , ,RTD
A, -SENSE, AUX V
20V PK MAX
20V PK MAX
mV Output
Black
Red
5520A calibrator
Phase meter
1 turn coil
or
5500A/coil
Tie Aux Com and
Normal Com together
using heavy gauge
wire or bus bar.
akp03f.eps
Figure 3. Phase Check Setup
mV Output Test
1.
Connect the equipment as shown in Figure 5.
2.
Set the calibrator output and coil for each step of Table 5.
3.
Verify that the voltmeter reading is within the “i1000S mV Output” limits of Table 5.