9150
User's Guide
11-2
4. Retrieve the original calibration errors from the unit or the Report of
Calibration.
5. Record the previous values for CE1, CE2, and CE3 and calculate new values
for CE1, CE2, and CE3 using the following formula.
Tmn –Tsn +CEn = CEm
Where:
Tmn is the temperature measured, Tsn (CTn) is the set-point temperature,
CEn is the old value for calibration error, and CEm is the new value for
calibration error.
6. Enter the new CEm value in the calibration parameter menu using either the
keypad or through the serial port.