GBK-50M GBK-110M
Instructions
6
Table 4. GBK-110M
3
2
4
5
6
7
8
9
10
1
11
12
Iez002.eps
No.
Description
No.
Description
Drive air supply regulator
G
High-pressure gas output gauge
Regulated booster drive pressure gauge
H
HIGH PRESSURE OUT Port
DRIVE port
I
Cover screw (2 locations)
D
Test gas supply pressure gauge
J
Booster drive air shut-off valve (booster ON/OFF)
E
SUPPLY port
K
Drive air supply filter (inside)
F
High-pressure gas output regulator
L
Drive air filter bowl drain valve (inside)
To install the Product, follow the steps below (see Tables 3 and 4):
1. Place the Product in the selected location.
2. Close both regulators, (
) and (
F
). Rotate the regulator counter-clockwise until no spring force is felt
Note
The drive adjust regulator (
) has a stop that prevents continued counter-clockwise rotation.
It also has a locking mechanism to prevent accidental adjustment. To unlock, pull the knob up.
3. Close the booster drive shutoff valve (
J
).
4. Connect shop drive air supply to the 1/4 in. NPT female DRIVE port with supplied tubing.
Содержание GBK-110M