FIBER OneShot
™
Length Meter
Users Manual
Страница 1: ...PN 3374501 October 2009 2009 Fluke Corporation All rights reserved Printed in USA All product names are trademarks of their respective companies FIBER OneShot Length Meter Users Manual...
Страница 2: ...ty on Fluke Networks behalf To obtain service during the warranty period contact your nearest Fluke Networks authorized service center to obtain return authorization information then send your defecti...
Страница 3: ...LIGHT 8 I O R index of refraction 8 ft m 8 dB LIMIT 8 The Connector Adapter 9 How to Clean Connectors and Adapters 9 How to Clean the Connector on the Meter 9 How to Clean Bulkhead Connectors sources...
Страница 4: ...ii FIBER OneShot Users Manual...
Страница 5: ...oting procedures and other support services To register online go to www flukenetworks com registration Or send the registration card that came with this product to Fluke Networks The Fluke Networks K...
Страница 6: ...o not use a magnifying device to look at the optical outputs without the correct filter Use of controls adjustments or procedures that are not in this manual can cause exposure to hazardous radiation...
Страница 7: ...cords can cause unreliable measurements Use the correct procedures to clean all fiber connectors before each test If you do not do this or if you use incorrect procedures you can get unreliable test...
Страница 8: ...ately 1500 tests before you must replace the batteries Physical Features See Figure 2 A Output port with SC adapter and UPC endface B Starts a test C On off key D Navigation keys E Press to see the se...
Страница 9: ...5 Physical Features fjy04 eps Figure 2 Physical Features fjy06 eps Figure 3 How to Identify an APC Connector Green housing Angled ferrule W Do not connect APC connectors to the meter...
Страница 10: ...l on the fiber CheckActive and ACTIVE LINE blink if there is an optical signal on the fiber The meter will not do a test if there is an optical signal on the fiber WCaution If ACTIVE LINE blinks immed...
Страница 11: ...e meter completes a length measurement G Shows as the meter measures length H J Shows when the length is more than the range of the meter See page 14 I Can show for short fibers when the meter cannot...
Страница 12: ...on is the ratio of the speed of light in a vacuum to the speed of light in a fiber The meter uses the index of refraction to calculate length If you increase n the calculated length decreases The defa...
Страница 13: ...de wipers or swabs to clean connectors You can purchase these supplies from Fluke Networks How to Clean the Connector on the Meter 1 Turn off the meter 2 Remove the connector adapter to get access to...
Страница 14: ...10 FIBER OneShot Users Manual fjy02 eps Figure 5 How to Remove and Install the Connector Adapter Unlock Lock Ferrule and fiber endface Remove...
Страница 15: ...for example the Fluke Networks FiberInspector Video Microscope to do an inspection on the connector How to Clean Connector Ends 1 Touch the tip of an optical grade solvent pen or swab soaked in solven...
Страница 16: ...the end of the fiber 1 Clean all fiber connectors 2 Connect the fiber to the meter as shown in Figure 6 To connect to APC connectors use a UPC APC patch cord To connect to UPC connectors use a UPC UPC...
Страница 17: ...eps Figure 6 How to Make Connections UPC connector only UPC connector only UPC UPC or APC APC connectors with an adapter UPC to UPC or APC to APC connection Note Always use patch cords that comply wi...
Страница 18: ...r instructions on how to clean the connectors The meter is connected to a PC connector PC connectors cause large reflections that the meter shows as the end of the fiber The meter shows J6000 m or J99...
Страница 19: ...arningX To prevent possible fire electrical shock personal injury or damage to the meter Do not open the case You cannot repair or replace parts in the case Use only replacement parts that are approve...
Страница 20: ...46 4519 Australia 61 2 8850 3333 or 61 3 9329 0244 Beijing 86 10 6512 3435 Brazil 11 3759 7600 Canada 1 800 363 5853 Europe 44 0 1923 281 300 Hong Kong 852 2721 3228 Japan 03 3434 0510 Korea 82 2 539...
Страница 21: ...tor Mini Video Microscope and Fiber Optic Cleaning Kit FT525 FT120 FiberViewer Microscope 200X FT120 FT140 FiberViewer Microscope 400X FT140 Fiber Optic Cleaning Kit with carrying case NFC Kit Case Fi...
Страница 22: ...22 2 No 61010 1 04 EN 61010 1 2nd Edition 2001 EN60825 1 2 2006 Altitude 3000 m EMC EN 61326 1 2004 Battery type 2 AA alkaline batteries no battery charger Battery life1 1500 tests Laser safety Class...
Страница 23: ...1 45 to 1 5 factory default is 1 468 Auto turn off Automatically turns off after 5 minutes if no keys are pressed Backlight turns off first Factory calibration interval None Output wavelengths 1310 nm...
Страница 24: ...3 dB above the noise floor 5 With a reflectance range from 40 dB to 14 dB 6 user configurable Index of Refraction IOR error the incident location error Incident location error for reflective incidents...
Страница 25: ...n Safety CSA C22 2 No 61010 1 04 EN 61010 1 2001 2nd Edition EN60825 1 2007 Class 1 Laser EMC EN 61326 1 EN61000 4 2 3 80 MHz to 2 7 GHz 3V m Telephony Requirements Class B Emissions P Conforms to rel...
Страница 26: ...22 FIBER OneShot Users Manual...