Set the Reference

The reference serves as the baseline power level for loss 
measurements.

Set the reference at these times:

At the beginning of each day using the remote end setup you 
will use that day. 

Anytime you reconnect a test reference cord to the module’s 
output or other source.

Anytime the tester warns you that the reference is out of date.

Anytime you see a negative loss measurement. (See the 
Technical Reference Handbook for more information.)

Anytime the tester tells you to set the reference.

To set the reference

1

Turn on the tester and smart remote and let them sit for 5 
minutes. Allow additional time if the modules have been 
stored above or below ambient temperature.

2

Turn the rotary switch to SETUP, then select Fiber Loss. Set the 
following on the Fiber Loss tabs (press 

C

to see other tabs):

Test Limit: Select the test limit required for the job. Press 

J

 More to see other lists of limits.

Fiber Type: Select the fiber type to be tested. 

Remote End Setup: Set to Smart Remote.

Bi-directional: Enable this if you are required to test the 
fiber in both directions.

Number of Adapters and Number of Splices: Enter the 
number of adapters and splices that will be added to each 
direction of the fiber path after the reference is set. 

Connector Type: Select the connector type used in the 
cabling to be tested. Select General if the exact type is not 
listed.

Test Method: Select 1 Jumper if you use the reference and 
test connection shown in this manual.

3

Turn the rotary switch to SPECIAL FUNCTIONS, then select Set 
Reference

4

The Set Reference screen shows reference connections for the 
test method you selected. Figure 2 shows Smart Remote mode. 
Clean and inspect the connectors on the testers and test 
reference cords, connect the tester and smart remote; then 
press 

P

Содержание DTX-CLT CertiFiber

Страница 1: ...t versions of the DTX CLT CertiFiber Users Manual and the DTX Series Technical Reference Handbook provided on the DTX CLT CertiFiber Product Manuals CD and on the Fluke Networks website WWarningX To avoid possible fire electric shock personal injury damage to equipment or inaccurate test results read the safety information in the DTX CLT CertiFiber Users Manual before using the tester A Power the ...

Страница 2: ...select Current Folder Select an existing folder or press J Create Folder to create a new folder Enter job information On the Instrument Settings menu press C to show the tab with the Operator Site and Company names To enter a new name select a setting press J Create then use the softkeys BC A D and H for editing Press N when you are finished C Clean and Inspect All Connectors WCaution When using t...

Страница 3: ...ress Cto see other tabs Test Limit Select the test limit required for the job Press J More to see other lists of limits Fiber Type Select the fiber type to be tested Remote End Setup Set to Smart Remote Bi directional Enable this if you are required to test the fiber in both directions Number of Adapters and Number of Splices Enter the number of adapters and splices that will be added to each dire...

Страница 4: ...tch to AUTOTEST then press P 4 Press H to see the test results The results for each fiber Input Fiber and Output Fiber must be no more than these limits WCaution An overall result of PASS does not show that the test reference cords are good You must compare the loss to the limits given here to make sure the cords are good Multimode test reference cord 850 nm 0 11 dB or less 1300 nm 0 10 dB or less...

Страница 5: ...utput Fiber When you have good test reference cords set the reference and do these steps again before you test a fiber link 6 Save the test results to show that you used good test reference cords to test the link Note Fluke Networks recommends that you use this procedure to make sure your test reference cords are good before you test each fiber link gfa03 eps Figure 3 Testing the Test Reference Co...

Страница 6: ...o a Test 1 Connect to the fiber you will test Figure 4 shows Smart Remote mode 2 Turn the rotary switch to AUTOTEST then press P gda04 eps Figure 4 Testing a Fiber Link in Smart Remote Mode 1 Jumper Method ...

Страница 7: ...te at www flukenetworks com registration Contacting Fluke Networks www flukenetworks com support flukenetworks com 1 425 446 4519 Australia 61 2 8850 3333 or 61 3 9329 0244 Beijing 86 10 6512 3435 Brazil 11 3759 7600 Canada 1 800 363 5853 Europe 31 0 40 2675 600 Hong Kong 852 2721 3228 Japan 03 6714 3117 Korea 82 2 539 6311 Singapore 65 6799 5566 Taiwan 886 2 227 83199 USA 1 800 283 5853 Visit our...

Страница 8: ...nty on Fluke Networks behalf To obtain service during the warranty period contact your nearest Fluke Networks authorized service center to obtain return authorization information then send your defective product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTICULAR PURPOSE ARE EXPRESSED OR IMPLIED FLUKE NETWO...

Отзывы: