DATAPAQ TP3
Specifications and Operation
13
Thermocouples
Available for a single thermocouple type, or for combinations of
up to three types: B, J, K, N, R, S, T (see specifications for each,
below).
Analog inputs:
Current
Voltage
See p. 29.
Measurement range 4–20 mA.
Accuracy ±0.1% at 24°C/75°F.
Measurement range 0–10 V.
Accuracy ±0.1% at 24°C/75°F.
Operating temperature
(of the logger itself)
Dependent on battery type:
Rechargeable −40°C to 70°C/−40°F to 158°F.
Alkaline
−40°C to 55°C/−40°F to 131°F.
Lithium
−40°C to 110°C/−40°F to 230°F.
Humidity range
0–100% non-condensing.
Operating pressure
(excluding limitations on
battery)
10
−7
bar at 20°C to 20 bar at 110°C
1.5
×
10
−6
psi at 68°F to 290 psi at 230°F
Real-time monitoring
Hardwired (serial) telemetry via communications lead (p. 55),
or via Bluetooth (p. 32), as standard.
Radio (RF) telemetry via optional built-in transmitter.
Data capacity
Over 3.6 million data-points in total, plus associated pre-trigger
data (see p. 54) and calibration data. This is sufficient for, e.g., 10
runs of 10 hrs with 10 probes and sample interval 5 sec. Using
fewer probes (see p. 43) increases the possible run-time.
Hot-data protection
By non-volatile memory, and software warning if reset attempted
before download.
Logger reset
By Insight (p. 41) or by start button (using previous reset
options) (p. 47).
Sample interval:
1
No telemetry
Hardwired telemetry
Radio telemetry
2
1–3 channels
0.1 s to 50 min.
4–9 channels
0.2 s to 50 min.
10 channels
0.3 s to 50 min.
11–20 channels 1 s to 50 min.
1–10 channels 1 s to 50 min.
11–20 channels 2 s to 50 min.
1–20 channels 2 s to 50 min.
In Japan
, 4 s to 50 min.
Data-collection start
No trigger, Start button, Date and time, Rising temperature,
Falling temperature.
Pre-trigger data stored
Yes (configurable; see p. 53).
Multiple runs
Collect data from up to 10 runs before downloading (see p. 45).
Multiple events
Up to 10 events (different probe selections, sample intervals and
trigger modes for different stages of a profile run; see p. 45).
Communications
USB 2.0, Mini-B socket.
Bluetooth (p. 29) (
not available in all countries
).
PC/software compatibility See p. 37.
cont. >>