Comark RF500A/AP Service Manual
Appendix B – Status Indicators
Appendix B – Status Indicators
The Status Indicators are found on the LHS of the Gateway and indicate self-diagnosed faults when accompanied by the
Font Panel Fault indicator.
O N / O F F
A U X
S T A T U S
1
2
3
4
R F 5 0 0 A
S E R I A L
N o .
C o m a r k I n s t r u m e n t s , 5 2 H u r r i c a n e w a y ,
N o r w i c h , N o r f o l k , N R 6 6 J B . U K
T e l : 0 1 6 0 3 ( + 4 4 1 6 0 3 ) 2 5 6 6 4 7
E m a i l : s e r v i c e @ c o m a r k l t d . c o m
w w w . c o m a r k l t d . c o m
C o m a r k I n s t r u m e n t s , P O B o x 9 0 9 0 ,
E v e r e t t , W A 9 8 2 0 6 , U S A
T e l : ( 5 0 3 ) 6 4 3 5 2 0 4
E m a i l : s e r v i c e @ c o m a r k U S A . c o m
w w w . c o m a r k U S A . c o m
R F 5 0 0 A P
M A D E I N U K
F O R I N D O O R U S E O N L Y
D O N O T M O V E W H E N
I N O P E R A T I O N
Figure 2 - Status Indicators
Fault Code
Status Indication
Fault Description
None
4
3
2
1
No Fault Detected
1
4
3
2
1
Backup Battery Voltage Fault.
2
4
3
2
1
Backup Battery Temperature Error
3
4
3
2
1
5V Power Supply Fault
4
4
3
2
1
3VM Power Supply Fault
5
4
3
2
1
Overheat Error
6
4
3
2
1
External Power Supply Fault
7
4
3
2
1
Processor Module Fault (3 successive failures)
Battery Pack Pre-Charge Indication
If on application of external power the backup battery does not have sufficient charge to run the Gateway should a
power fail condition arise then Gateway will not start but instead starts a pre-charge mode. This is indicated by the
following repeating pattern in the Status Indicators.
4
3
2
1
4
3
2
1
4
3
2
1
4
3
2
1
4
3
2
1
4
3
2
1
4
3
2
1
This mode lasts for a duration of 20 minutes minimum.
17 November 2011
Page 52 of 65