CO-205 Instruction Sheet

Trim and fold to 4.125 in (W)  x 8.25 (H)

Page 

2

Using the Aspirator

To use the aspirator:

1.  Snap the aspirator cap to the CO-210 or CO-220.

2. 

Downstream from the furnace but before any intersections
to the flue, make a 5/16” sampling hole in the flue to
accommodate the aspirator probe.

3.  Insert about half of the aspirator probe into the flue.

 

Caution: Do not leave the aspirator probe in the flue for
extended periods of time (longer than 1 hour).

4.  Turn on the CO-210 or the CO-220. (If the CO-210 is

connected to a meter, turn on and configure the meter.)

5.  Squeeze the aspirator bulb 4 to 5 times or until the

CO-210 or CO-220 shows a constant reading.

6.  After sampling is complete, plug the sampling hole using

material appropriate for your flue.

1

5

3

2

4

XKOO1F.EPS

A

Aspirator probe assembly (Fluke P/N 691565)

B

PTFE filter (Fluke P/N 691188. Whatman 7590-002)

C

Aspirator hose (Viton)

D

Aspirator cap (Fluke P/N 691185)

E

Fluke CO-210 or CO-220 (Not included in 205 Aspirator Kit)

Figure 1. CO-205 Aspirator Kit

Specifications

Hose Dimensions:

Inside diameter
Outside diameter
Length

3.175 mm (1/8 inch)
6.35 mm (1/4 inch)
122 cm (48 inches)

Filter Pore Size:

1 µm

Operating Temperature:

Probe:
Hose:

371 

C (700 

F)

204 

C (400 

F)

Storage Temperature:

-30 

C to +60 

C (-22 

F to +140 

F)

Replacing the Filter

The PTFE filter on the aspirator hose can clog with use.  A
clogged filter reduces the accuracy of the CO reading by
impeding air flow to the measuring device. Examine the filter
periodically and replace as necessary.

If you notice water in the hose on 

the instrument side of the

filter, the filter has become saturated. Remove the filter, drain
the water, and replace the filter.

Warranty

Limited Warranty & Limitation Of Liability

This Fluke product will be free from defects in material and workmanship for
one year from the date of purchase. This warranty does not cover fuses,
disposable batteries or damage from accident, neglect, misuse or abnormal
conditions of operation or handling. Resellers are not authorized to extend any
other warranty on Fluke’s behalf. To obtain service during the warranty period,
send your defective tester to the nearest Fluke Authorized Service Center with
a description of the problem.

THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES,
SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED
OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT,
INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING
FROM ANY CAUSE OR THEORY. Since some states or countries do not
allow the exclusion or limitation of an implied warranty or of incidental or
consequential damages, this limitation of liability may not apply to you.

Fluke Corporation

Fluke Europe B.V.

P.O. Box 9090 

P.O. Box 1186

Everett, WA.

5602 B.D. Eindhoven

98206-9090

The Netherlands

10/96

Отзывы: