Product Overview and Specifications
Specifications
1
1-11
Specifications
General Specifications
Fuse Protection for Resistance
...................... 0.44 A (44/100 A, 440 mA), 1000 V FAST Fuse, Fluke specified part
only
Power Supply
Battery power ...................................................... BP500 smart battery pack: double cell lithium-ion, 7.4 V, 3000 mAh
Battery life ........................................................... >8 hours in continuous full-load operation
Battery charging time ..........................................
≤
4 hours
Power adapter output voltage ............................. Use only BC500 battery charger: 18 V, 840 mA
Line power .......................................................... 100 V ac to 240 V ac adapter with country specific plug
Frequency ........................................................... 50 Hz to 60 Hz
Temperature
Operating ............................................................ 0 °C to 40 °C
Storage ............................................................... -20 °C to 50 °C
Lithium-ion battery charging ............................... 0 °C to 40 °C
Relative Humidity (non-condensing, 10 °C)
Operating ............................................................
≤
80 % at 10 °C to 30 °C
≤
75 % at 30 °C to 40 °C
Altitude
Operating ............................................................ 2,000 m
Storage ............................................................... 12,000 m
Temperature Coefficient
....................................... 0.1 x (specified accuracy) /°C (<18 °C or >28 °C)
Size
......................................................................... 58 x 103 x 220 (mm)
Weight
.................................................................... 850 g
Memory
Data/Setup flash memory ................................... 4 MB
Real-Time Clock
.................................................... Time and date stamp for measurement. The RTC works >50 days
without battery.
IP Rating
................................................................ IEC 60529: IP 40
Safety
..................................................................... IEC 61010-1, IEC 61010-2-030, IEC 61010-031, Pollution Degree 2
600 V CAT III; Derated to CAT II with CAT II probe cap installed
EMI, RFI, EMC, RF
................................................. EC 61326-1: Portable Electromagnetic Environment; IEC 61326-2-2
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses
conductively-coupled radio frequency energy that is necessary for
the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other
than domestic and those directly connected to a low-voltage
power supply network that supplies buildings used for domestic
purposes.
EN 300 328, EN 301 489-1, EN 301 489-17, FCC Part 15 Subpart C
Sections 15.207, 15.209, 15.249
CONTAINS FCC IDs: T68-FWCS, XDULE40-S2
IC: 6627A-FWCS, 8456A-LE4S2
Electromagnetic Compatibility
............................ Applies to use in Korea only. Class A Equipment (Industrial
Broadcasting & Communication Equipment)
[1]
[1]
This product meets requirements for industrial (Class A) electromagnetic wave equipment and seller or user should take notice of it.
This equipment is intended for use in business environments and is not to be used in homes.
Содержание BT510
Страница 6: ...BT520 Users Manual iv...
Страница 8: ...BT520 Users Manual vi...
Страница 10: ...BT520 Users Manual viii...
Страница 14: ...BT520 Users Manual 1 4 6 7 11 12 13 16 8 10 14 15 9 1 3 2 4 5 hsn056 eps Figure 1 1 Standard Equipment...
Страница 24: ...BT520 Users Manual 1 14...
Страница 48: ...BT520 Users Manual 3 18...
Страница 66: ...BT520 Users Manual 7 6 hsn032 eps Figure 7 3 Charge the Battery...
Страница 68: ...BT520 Users Manual 7 8...