Infusion Pump Analyzer
Operating Instructions
3
3-19
Occlusion Test Screen
This is the screen that allows the operator to control the
occlusion test. It has a menu across the bottom of the
screen that changes to reflect the progress of the test. It
is not necessary to prime the transducer to perform an
occlusion test.
The operator sets the units of pressure displayed on this
screen from the
TEST PARAMETERS SCREEN.
Occlusion Test Screen (Wait Mode)
When the
OCCLUSION TEST
screen is first entered it
puts the transducer into a Tare measuring mode (i.e.
subtracts the ambient pressure). While the transducer is
in this mode the
WAIT
prompt is displayed.
Wait
This menu option is never active, it informs the operator
to wait until the transducer has taken a tare reading.
When the tare reading has completed this, the menu
option changes to
START
, and the instrument advances
to the start mode. The pump should be stopped or
running very slowly during the wait period.
Status
This invokes the
STATUS ALL CHANNELS
screen.
ESC
Selecting
ESC
or pressing the
ESC
key on the external
keyboard takes the operator back to the
CHANNEL
SETUP
menu.
C H A N N E L 1 O C C L U S I O N
0.00
PRESSURE
psi
TIME
00 : 00 : 00
Peak Pressure
psi
Time of Peak 00: 00: 00
WAIT
STATUS ESC
Содержание Biomedical IDA-4 Plus Operators
Страница 1: ...IDA 4 Plus Infusion Pump Analyzer Operators Manual...
Страница 2: ......
Страница 4: ...IDA 4 Plus Operators Manual ii...
Страница 8: ...IDA 4 Plus Operators Manual vi...
Страница 21: ...Infusion Pump Analyzer An Overview 2 2 5 Rear Panel Layout Drawing...
Страница 22: ...IDA 4 Plus Operators Manual 2 6...
Страница 24: ...IDA 4 Plus Operators Manual 3 2 Inlet Hose Connections Drawing...
Страница 68: ...IDA 4 Plus Operators Manual A 8...