4
Parts and Controls
The 9116A consists of a control panel, furnace core, and a back panel. Each
part and control is described below.
4.1
Control Panel
The controls to the furnace are located on
panels to the right of the instrument. The
upper portion of the panel is sloped and
contains the controller which is regularly
used during operation of the furnace. An
LED on the main display indicates when
the controller is sending power to the
heater. Red indicates the heater is on, green
indicates the heater is off, pulsing means
the heater power is being regulated by the
controller between 0 and 100%.
4.1.1
Controller
The controller has overall control of the
furnace. This sloped panel is located on the
upper right portion of the furnace (see
Fig-
ure 4
). The controller itself is a hybrid ana-
log/digital device utilizing the high stability
of analog circuitry with the flexibility of a
micro-processor interface and digital
controls.
The following controls and indicators are
present on the primary controller panel: (1)
the digital LED display, (2) the control but-
tons, and (3) control indicator LED.
(1) The digital display shows the set and
actual temperatures as well as various
other functions, settings and constants.
The temperature can be set in scale
units of either °C or °F.
(2) The control buttons (SET, DOWN, UP,
and EXIT) are used to set the furnace
temperature set-point, access and set
other operating and calibration
parameters.
17
POWER
O
I
SET
DOWN
UP
EXIT
9116A
FREEZE POINT FURNACE 550°C to 1000°C
419.03 C
HEATER
MODE
Figure 4.
Front Control Panel (cover
door remove)
Содержание 9116A
Страница 1: ...9116A Furnace User s Guide Rev 842101 Hart Scientific ...
Страница 24: ...19 Figure 5 Back Panel ...
Страница 28: ...9116A Furnace User s Guide 24 Figure 6 Controller Flow Chart ...
Страница 56: ...53 8 Fixed Point Cell Installation Instructions Procedure Figure 14 Place insulation in re entrant well ...