9102S
User’s Guide
20
ALPHA
This probe parameter refers to the average sensitivity of the probe between 0 and 100 °C. The value of
this parameter is set at the factory for best Product accuracy.
aLpha
Flashes for one second and then the ALPHA setting displays
38530
Current ALPHA setting
To change the ALPHA setting, press
u
or
d
.
38600
New ALPHA setting
To accept the new setting, press
SET
. Press
EXIT
to continue without storing the new value.
S
Accept the new ALPHA setting
DELTA
This probe parameter characterizes the curvature of the resistance-temperature relationship of the
sensor. The value of this parameter is set at the factory for best Product accuracy.
DeLta
Flashes for one second and then the DELTA setting displays
0.0000
Current DELTA setting
To change the DELTA setting, press
u
or
d
0.1000
New DELTA setting
To accept the new setting, press
SET
. Press
EXIT
to continue without storing the new value.
S
Accept the new DELTA setting
Digital Communication Interface
This Product can communicate with and be controlled by other equipment through the digital serial
interface.
With a digital interface, the Product may be connected to a computer or other equipment. This allows
the user to set the set-point temperature, monitor the temperature, and access any of the other
controller functions, all using remote communications equipment. Communications commands are
summarized in Table