Return the AC Calibrator to OPER. The DMM should 

read between 28.3 and 31.8V dc 
6.

 

Return the UUT to the scope output and set the level 

sine to 1V rms 160 MHz. The DMM should read between 0.9 
and 1.12V dc. Increase the frequency to 450 MHZ. The DMM 
should read between 0.70 and 1.41V dc. 
Troubleshooting 
Isolate a problem to the probe by verifying the operation of 
the probe’s input source and output DMM. Most probe 
malfunctions are caused by the failure of diodes CR1 or 
CR2. Suspect CR1 if there is no output when an input is 
applied to the probe. Excessive linearity error may be due to 
the failure of CR2. Use a continuity test for in-circuit diode 
testing. If the diodes are good, check the input and output 
connections to the printed circuit board  and the continuity of 
the output cable. 

CAUTION 

The diodes have glass bodies; care should be taken to avoid 
stressing them, particularly when preparing the leads. The diodes 
are also subject to failure from overheating. When soldering, hold 
the lead near the diode body with a pair of needle nose pliers. 
Avoid excessive heat. 

Probe Disassembly 
Use the following procedure to access the probe’s internal 
components, including the calibration adjustment R2: 
1.

 

Unscrew the bend relief and slide it away from the 

probe. 
2.

 

Lift off the probe cover. R2 is exposed. 

3.

 

Remove the probe tip by carefully grasping the tip with a 

pair of pliers and drawing it out of the front of the probe. 
4.

 

Unscrew the ground lead from the probe body. 

5.

 

Lift the rear of the printed circuit board slightly and pull it 

firmly away from the probe tip. 
6.

 

To reassemble the probe, logically reverse this 

procedure. Make sure that the gold pin on the front of the 
printed circuit board centers properly in the tip and that the 
ground-lead nut is aligned with the hole in the probe body. 

Cleaning 

Use a soft cloth dampened in a mild solution of detergent 
and water to clean the 85RF. Do not use solvents. 

85RF PARTS LIST 

 DESCRIPTION 

PART 

NO. 

C1 

Cap, Ceramic Chip, 220 pF 

758078 

CR1 Diode, 

Germanium 

629261 

CR2 Diode, 

Silicon 

639922 

R1 Resistor, 

cf, 

220k 

348953 

R2 Resistor, 

Var, 

500k 

574624 

R3 Resistor, 

cf, 

1.3M 

442558 

 Body, 

Probe 

606988 

 Cover, 

Probe 

581751 

 Tip, 

Probe 

607267 

 Wire, 

Ground 

Assy 

607291 

 Decal, 

Probe 

585842 

 Contact, 

Ground 

586560 

 Contact, 

Twist 

Pin 

294991 

 Tip, 

Insulated 

583682 

 Tip, 

IC 

583674 

 

www.

.com

[email protected]

1.800.561.8187

Отзывы: