Specifications 

AC to DC ratio:

 1:1

 

Ratio accuracy: (at 25 MHz and loaded with 10 M

e

 ) 

Above 0.5V: ± 0.5 dB 
Below 0.5V: ± 1.0 dB 
Add ±0.2 dB in RF fields of 1 V/m to 3 V/m 

Frequency response: (Relative to 25 MHz) 

100 kHz to 80 MHz ±0.5 dB 
*80 MHz to 200 MHz ±2.0 dB 
*200 MHz to 300 MHz ±3.5 dB 
*300 MHz to 400 MHz ±4.5 dB 
*400 MHz to 500 MHz ±5.5 dB 
*Referred to high and low inputs at Probe tip. 

Extended frequency response: 

Useful for relative measurements from  
20 kHz to 700 MHz. 

Response: 

Responds to the peak value of an input and is calibrated 
to read rms value of a sine wave. 

Voltage range:

 0.25 V rms to 30V rms

 

Maximum input voltage:

 30 V rms, 200 V dc

 

Input capacitance: 

Approximately 9 pF

 

Temperature range: 

Operating: +10 °C to +35 °C 
Storage: -40 °C to +75 °C 

Humidity: 

<90 % R.H.

 

Output connector:  

Fits standard 0.75 inch dual banana connectors 

Accessory:  

BNC to Probe Adapter, P/N 574756 

Pollution degree: 

2

 

Operating altitude: 

3000 m

  

Warranty: 

1 year

  

Voltmeter Compatibility 

The 85RF II can be used with a dc voltmeter that has an 
input resistance of 10 M

e

, ±10 %. Voltmeters with higher 

input resistances are usable if an external shunt is put 
across the input terminals. The shunt value is set to 
decrease the input resistance to 10 M

e

 
 
 
 

Содержание 85RF II

Страница 1: ... Product Look for cracks or missing plastic Carefully look at the insulation around the terminals Limit operation to the specified measurement category voltage or amperage ratings Do not touch voltages 30 V ac rms 42 V ac peak or 60 V dc Do not use the Product around explosive gas vapor or in damp or wet environments Do not use and disable the Product if it is damaged Use the Product only as speci...

Страница 2: ...ue of an input and is calibrated to read rms value of a sine wave Voltage range 0 25 V rms to 30V rms Maximum input voltage 30 V rms 200 V dc Input capacitance Approximately 9 pF Temperature range Operating 10 C to 35 C Storage 40 C to 75 C Humidity 90 R H Output connector Fits standard 0 75 inch dual banana connectors Accessory BNC to Probe Adapter P N 574756 Pollution degree 2 Operating altitude...

Страница 3: ...the Probe 2 Touch the Probe tip to the circuit under test Frequencies greater than 100 MHz To measure frequencies greater than 100 MHz 1 Ground the 4 inch clip lead supplied with the Probe 2 Touch the Probe tip to the circuit under test Note Measurements above 100 MHz must be made with minimum cable lengths and a terminated 50 e system One inch or less of non terminated coaxial cable such as a BNC...

Страница 4: ...bration The performance and calibration test is a verification test to make sure that the 85 RF II is functional and correctly calibrated Use the procedure as an acceptance test and or calibration procedure the Probe has one adjustment which is R2 The recommended calibration interval is one year Note If the Probe cannot be calibrated to the specifications in this document the Probe could be damage...

Страница 5: ...e limits disassemble the Probe refer to the Probe Disassembly Section in this Instruction Sheet and adjust R2 for a DMM measurement between 0 990 V dc and 1 010 V dc Reassemble the Probe 3 Move the unit in test to the Normal output of the calibrator and decrease the output level to 0 250 V rms and 1 MHz The DMM will read between 0 223 V dc and 0 280 V dc 4 Set the AC Calibrator to STDBY decrease t...

Страница 6: ...of the printed circuit board up inch and pull it away from the Probe tip 6 To reassemble the Probe do the steps in this procedure backwards start with step five and end with step one Make sure that the gold pin on the front of the printed circuit board aligns correctly in the Probe tip and that the ground lead nut is aligned with the hole in the Probe body How to Clean the Probe WXWarning For safe...

Отзывы: